TIPS - QUESTIONS - ANSWERS
ILLUMINATION AND PROBE
WORK WELL BUT THE IMAGE IS
MISSING
THE IMAGE IS NOT CLEARLY
VISIBLE
THE IMAGE IS FLIPPED,
MIRRORED OR ROTATED
IMAGE MANIPULATION IS
CAUSING SMEARING
THE IMAGE COLOR IS
NOT CORRECT
THE SYSTEM DOES NOT RESPOND
A
The optic / light output of the probe
may be contaminated by moisture or
dirt. Clean the optics and / or side view
prism
B
The light source has not been started
or is displayed as OFF in the operation
menu.
C
Set the focus using the RF slide bar
l
3.2.6 LCD Control unit, page 21
D
Illumination is set manually and low. Set
the illumination slide bar to AUTO
E
Illumination is insufficient, apply addi-
nional image exposure time
F
Image exposure time is turned ON cau-
sing smearing images when moving the
probe. Turn OFF the additional image
exposure time.
A
Check if the video IN / OUT is set to INT
B
Check if the probe connectors are
clean and properly connected to the
base unit.
C
Is the probe tip damaged?
A
Perform a white balance calibration
B
Check the color calibration
C
Check the brightness calibration
A
Press RESET to reset all image manipu-
lations
A
If your system does not respond, turn
OFF the main switch and re-start the
system.
A
Press RESET to reset all image manipu-
lations