SuperPump AR System User Manual
4.4
Select Waveform
Press the Menu Selector Dial.
Turn the Menu dial until “1 Waveform Source” appears as shown
below.
0
0.0 % Gain
0.0 ml/Stroke
1 Waveform Source
While the message Waveform Source is on the screen, press the dial again. The screen will
now indicate the most recently used waveform source.
0
0.0 % Gain
0.0 ml/Stroke
Wfm: External
If the current waveform source is desired press the dial again to escape the waveform selector.
If a stored waveform is desired turn the dial clockwise and press the dial to select the displayed
waveform.
Refer to section 4.7 for more details on waveform options.
0
0.0 % Gain
0.0 ml/Stroke
Wfm: PHYSIO_70
4.5
Sync Pulse Delay
The Sync Pulse is a +5V signal with 10ms duration which can be used for synchronizing and
triggering equipment such as high speed cameras.
Press the Menu Selector Dial. Turn the menu dial until
“2 Sync Pulse Delay” is on the screen,
press the dial again. The screen will now indicate the current sync delay in msec.
0
0.0 % Gain
0.0 ml/Stroke
2 Sync Pulse Delay
To adjust the delay turn the dial until the correct delay is displayed and then press the dial. The
Sync Pulse Delay may be adjusted from 0 to 2000ms. A 0ms delay will trigger as the piston
crosses forward across the home position.