background image

Display Screen

•  

The compressor pump always shows the air pressure 

(mmHg) of the mattress on the display screen. 

It can also 

display the patient’s weight in Pounds when adjusting the 
“Hard or Soft” settings.

Power

•  When pushing the power button for first time, the mattress 
will begin inflating. This process takes approx. 50 minutes. 
When fully inflated, the mattress will begin alternating 
automatically. 

Hard Mode

• Increases the weight capacity in Pounds. 

Tapping the 

button each time will increase the weight by increments of 
5 Lbs. To increase faster, hold down the button.

Soft Mode

•  Decreases the weight capacity in Pounds. 

Tapping the 

button each time will decrease the weight by increments 
of 5 Lbs. To decrease faster, hold down the button.

Static Mode

•  Makes the mattress stop alternating and all cells will have 
equal pressure. User can adjust the firmness by pressing 
the Firm or Soft Key. The display will show the corresponding 
mmHg / Lbs. 

Seat Inflation Mode

• This allows the patient to set the bed in a seated position.
A “S” 

will show on the display screen. 

Lock/Unlock Mode

•  

Hold the “Lock/Unlock Mode” for 3 seconds to unlock 

panel and adjust settings. 

After 2 minutes of not using the 

buttons, the panel will lock automatically. 

OPERATING FUNCTIONS

Display Screen (mmHg or Lbs)

Power

Hard Mode (Lbs)

Static Mode

Seating Mode (60mmHg)

Lock / Unlock Mode

Soft Mode (Lbs)

Firm

Soft

Power

Seat Inflation Mode

Lock / Unlock

Static Mode

mmHg / Lbs

Automatic Pressure Monitor

mmHg / Lbs