5. To prolong battery life, press the On/Off Button to turn unit off after
testing is complete. If no action is taken, the unit will automatically
shut off after around 10 minutes.
BATTERY REPLACEMENT
1. Replace battery when “ ” appears in the lower right corner
of LCD display.
2. Pull battery cover off as shown in Figure 2.
3. Use pointed object such as pen to remove old battery. Discard
batter lawfully. Replace with new 1.5 DC button type LR41 or SR41,
UCC392, or equivalent. Be sure battery is installed with “+” polarity
facing up. (See figure 3,4)
4. Slide battery chamber back into place and attach cover.
CLEANING & DISINFECTING
The best way to clean the thermometer tip is by applying a
disinfectant (e.g. 70% ethyl alcohol) with a damp cloth. The
waterproof thermometer can be immersed in liquid or lukewarm
water for thorough cleaning and disinfection.
PRODUCT DISPOSAL
Please ensure environmental protection. Batteries do not
belong in the domestic waste. Please hand them in at
collection point or the municipal recycle material center as
special waste.
This symbol on products and/or accompanying documents
means that consumed electronic products must not be mixed with
conventional domestic waste.
Take these products to the corresponding collection points for correct
treatment and recycling, where they will be accepted free of charge.
For more information on the closest collection point, Please enquire
with your local authorities.
LEGAL REQUIREMENTS & GUIDELINES
This thermometer is manufactured according to the system of
EN12470-3:2000+A1:2009, ASTME1112-00. This thermometer complies
with Part 15 of the FCC Rules. Operation is subject to the following two
conditions: (1) this thermometer may not cause harmful interference,
and (2) this thermometer must accept any interference received
including interference that may cause undesired operation. And the
manufacturer is ISO 9001: 2008 and ISO 13485:2003 + AC: 2007
certified.
SPECIFICATIONS
• Waterproof and flexible features for using conveniently
• Range: 32.0°C – 42.9°C (90.0°C – 109.9°F)
• Accuracy: ±0.1 °C, 35.5°C –42.0°C (±0.2°F, 95.9°F-107.6°F)
±0.2°C under 3.5°C or over 42.0°C
(±0.4°F under 95.9°F or over 107.6°F)
at standard room temperature of 18°C to 28°C
(64.4°F to 82.4°F)
• Display: Liquid crystal display, 3 1/2 digits
• Memory: For storing the last measured value
• Battery: One 1.5V DC. Button battery (size LR41 OR SR41, UCC392)
• Battery Life: Approx 200 hours
• Dimension: 13.9cm x 2.2cm x 1.3 cm (L x W x H)
• Weight: Approx 11.0 grams including battery
• Using Environment
Temperature: 5°C ~ 40°C (41°F ~ 104°F)
Relatively humidity: <85%RH
• Storage and transportation condition
Temperature: -10°C ~ 55°C (14°F ~ 131°F)
Relatively humidity: <95%RH)
• Waterproof level: Waterproof
• Classification: TYPE B
ACTIVATE YOUR GUARANTEE
This product is guaranteed to be free from manufacturing defects.
We will replace or exchange any product that is defective to your
complete satisfaction. This warranty extends one year from the
purchase date, as long as the product has not been misused or
abused.
© Copyright and Distributed by Vive Health
8955 Fontana Del Sol Way, Naples, FL 34109
All Rights Reserved. Made in China.
Email: [email protected]
Figure 2
Figure 3
Figure 4
To ensure your warranty,
please register your product at
www.vivehealth.com/register.
Product Code: DMD1024GRY