105 PRINT R$,I
120 ! Issue test and command to run.
130 ! OUTPUT @V944;"ADD,VDC,1000,100,1.234e-05,0,10;RUN;"
140 ! Poll until test is complete.
150 Stb=SPOLL(@V944)
160 IF Stb<8 THEN 150
170 PRINT "Stb=",Stb
180 OUTPUT @V944;"RSLT?;"
190 ENTER @V944;R$
200 PRINT R$
210 STOP
220 END
ADD, {MEG_OHM}
This command loads a single step test into the 944i's test buffer. The added test may
then be run with the RUN command.
Arguments:
<MEG_OHM>,
Selects MEG_OHM function
<test voltage>,
100 - 5000 V
<slew rate>,
1 - 1999 V/S
<min resistance>, 100000
Ω
<test time>;
1 - 999 S
Example
:
10 ! File: T3.IBW
20 ! This test sends a MEG_OHM command to the 944i, waits
30 ! for test completion, and then returns the result
40 ! to the bus controller.
50 DIM Result$[120]
60 ASSIGN @V944 TO 1410
70 REMOTE @V944
80 ABORT 14
90 CLEAR 14
100 ! Issue test and command to run.
110 OUTPUT @V944;"ADD,MEG_OHM,1000,100,1.0e+6,5;RUN;"
120 ! Poll until test is complete.
130 Stb=SPOLL(@V944)
140 IF Stb<8 THEN 130
150 PRINT "Stb=",Stb
160 ! Get test results.
170 OUTPUT @V944;"RSLT?;"
180 ENTER @V944;Result$
190 PRINT Result$
200 PRINT
210 BEEP
220 STOP
230
END
ADD, {LO_OHM}
This command loads a single step test into the 944i's test buffer. The test may then be run with the RUN command.
Arguments for
944i
measurement:
<LO_OHM>,
Selects
LO_OHM
function
<max resistance>, 0.001 - 8.5 Ohms (0.001 – 500 with LO-500 option)
<min resistance>, 0 - 8.5 Ohms (0 – 499.999 with LO-500 option)
<test time>;
1 - 999 S