With pipettes of a nominal volume 

≤ 

50 µl, the error limits are usually smaller than 0.5 

µl

. Due to

this small error limit, the evaporation of water during the test procedure has a relatively large influ-
ence on the result. Therefore, the testing of pipettes of this size requires a test procedure which 
largely prevents evaporation. To provide a sufficient evaporation protection, special pipette calibration
balances including an evaporation trap can be used.

Pipettes with nominal volume  

≤ 

50 µl

Through the use of calibrated testing devices (balance and thermometer), the requirement of DIN EN
ISO 9001, DIN EN ISO 10012 and DIN EN ISO / IEC 17025 to refer the test to national standards is ful-

filled. The calibration of the balance e.g. can be carried out either by DAkkS calibration or official cert-
ification of the balance, or by calibrating the balance with appropriate weights that are traced to the 
national standard. The calibration of the thermometer, hygrometer and barometer can also be carried 
out by DAkkS calibration or official certification, or by a comparison with thermometers that are traced  

Traceability of test results to national standards

to the national standard (under defined conditions).

5. Calibration - Checking the volume

 1. Set the nominal volume at 10% or 20%.

Enter temperature into test record.

 

2. Determine temperature of the test liquid. 

3. Place the weighing vessel filled with test liquid on the balance and tare balance.

Note: For multichannel pipettes, each channel has to be tested seperately!

 

4. Attach new pipette tip. For multichannel pipettes, attach a new pipette tip only to one channel.

Perform conditioning of the pipette. Therefore, aspirate and release test liquid 5 times.

The conditioning increases the accuracy of the test.

 

5. Aspirate test liquid from the recipient vessel. Push the pipetting button to the first stop. Immerse

the pipette tip vertically 2 - 4 mm into the test liquid. For 5 ml and 10 ml pipettes, immerse the
tip 3 - 6 mm. Then release the pipetting button steadily and smoothly.

Observe waiting time: Remain approx. 1 second in the test liquid. For 5 ml and 10 ml pipettes,

wait approx. 10 seconds.

 

6. Release the test liquid into the weighing vessel. Lean pipette tip in an angle of 30° - 45° against the

wall of vessel. Push the pipetting button to the first stop and hold. Push to the second stop to empty
the pipette tip completely. Wipe off pipette tip against the wall of the vessel over a distance of
approx. 10 mm. Then release the pipetting button smoothly.

Содержание micropipette

Страница 1: ...Standard Operating Procedure SOP VITLAB micropipette VITLAB GmbH Linus Pauling Str 1 63762 Grossostheim Germany tel 49 6026 97799 0 fax 49 6026 97799 30 info vitlab com www vitlab com...

Страница 2: ...Table of contents 1 Introduction 3 2 Preparation of the pipette for testing 4 3 Test instruments and environment 6 4 Functional testing 7 5 Calibration Checking the volume 8 6 Evaluation 9 7 Tables 1...

Страница 3: ...01 DIN EN ISO 10012 and DIN EN ISO IEC 17025 Owing to its 8 respectively 12 channels the VITLAB micropipette 8 12 can carry out 8 resp 12 pipettings in one action However this advantage results in an...

Страница 4: ...ready VITLAB micropipette Use only manufacturerer s original parts Pipette tips Use only appropriate tips Optimum results can only be obtained with VITLAB pipette tips 2 3 Cleaning Clean pipette shaf...

Страница 5: ...ing manual Error Possible cause Measures Pipette tip is no longer tight Scratches on the pipette shaft tip Obtain spare parts see operating manual Device is stiff and or leaking Seal Piston soiled or...

Страница 6: ...ent error of 0 2 C Hygrometer Considering the measuring tolerance of the hygrometer a relative atmospheric humidity of 40 60 should be reached Balance Recommended specifications see table Selected vol...

Страница 7: ...wall of the vessel and wipe off against the wall jerk free Eject the tip Enter results in the test record Possible errors and the resulting measures For other measures and more troubleshooting advice...

Страница 8: ...standards to the national standard under defined conditions 5 Calibration Checking the volume 1 Set the nominal volume at 10 or 20 Enter temperature into test record 2 Determine temperature of the tes...

Страница 9: ...8 values for the 12 channel pipette 6 Evaluation The obtained weighing values from the gravimetric test are only the mass values of the pipetted volume without the correction of the air buoyancy To ob...

Страница 10: ...n the operating manual Recalibrate the VITLAB micropipette according to the steps in the operating manual If these measures are not successful we recommend you to make use of the VITLAB calibration se...

Страница 11: ...tor z ml g Temperature C 1 0020 1 0020 23 23 5 1 0035 1 0036 16 16 5 1 0021 1 0022 24 24 5 1 0038 1 0039 17 17 5 1 0023 1 0024 25 25 5 1 0040 1 0041 18 18 5 1 0025 1 0026 26 26 5 1 0043 1 0044 19 19 5...

Страница 12: ...6 10 5 1 20 10 2 100 50 10 200 100 20 1000 500 100 5000 2500 500 10000 5000 1000 1 2 0 8 1 0 0 5 0 8 0 3 0 8 0 3 0 8 0 3 0 8 0 3 0 6 0 3 For calibration the error limits to be observed by the operator...

Страница 13: ...1 2 3 0 0 3 0 6 1 5 10 5 1 50 25 5 100 50 10 200 100 20 300 150 30 2 4 1 6 2 0 0 8 1 6 0 6 1 6 0 6 1 6 0 6 For calibration the error limits to be observed by the operator must individually defined by...

Страница 14: ...ominal volume Serial number Customers identification 5 Weighing values and results of the gravimetric test VITLAB continuous E RS VITLAB genius VITLAB simplex VITLAB TA VITLAB micropipette Weighing va...

Страница 15: ...st 35 Correction factor Z 1 Instrument Type VITLAB micropipette 8 12 8 channel 12 channel Nominal volume Serial number Customers identification Channel No Weighing value 1 2 3 4 5 6 7 8 9 10 11 12 X1...

Страница 16: ...inal volume mg A nominal CV nominal Channel No Weighing values 1 2 3 4 5 6 7 8 9 10 11 12 X1 mg X2 mg X3 mg X4 mg X5 mg X6 mg X7 mg X8 mg X9 mg X10 mg X mean mg V mean l A found CV found A nominal CV...

Страница 17: ...nal Channel No Weighing values 1 2 3 4 5 6 7 8 9 10 11 12 X1 mg X2 mg X3 mg X4 mg X5 mg X6 mg X7 mg X8 mg X9 mg X10 mg X mean mg V mean l A found CV found A nominal CV nominal Result A Result CV The t...

Страница 18: ...e been carefully cleaned and decontaminated before shipment That the instruments pose no danger through bacteriological chemical radiological or viral contamination That she he is aware that shipment...

Отзывы: