Guidance and manufacturer’s declaration – electromagnetic emissions
The 4000 Respiratory Monitor lung monitor is intended for use in the electromagnetic
environment specified below. The customer or the user of 4000 Respiratory Monitor lung monitor
should assure that it is used in such an environment.
Emissions
test
Compliance
Electromagnetic environment
–
guidance
RF emissions
CISPR 11
Group 1
The 4000 Respiratory Monitor lung monitor uses RF energy
only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The 4000 Respiratory Monitor lung monitor is suitable for use
in all establishments, including domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
Harmonic
emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/
flicker
emissions
IEC 61000-3-3
Class A