47
Problem
Possible cause
Fix
The control
unit cannot be
switched on.
The timer function is
activated but no period
of operation is set.
Check the position of the main
switch. If it is in position
II,
the
timer function is activated. In
order to start the control unit you
have to set the rotary knob of the
timer to the required period of
operation.
If you do not wish to use the timer
function, press the main switch to
position
I.
The control unit is
not connected to the
mains.
Insert the mains plug into the
socket.
There is no mains
voltage.
• Try another mains socket.
• Check the fuse of your mains
connection.
The fuse in the control
unit has been tripped
due to overvoltage
(e. g. lightning strike).
Contact the customer service
department.
The cuffs open
by themselves
during use.
In the event of over-
load, the zips on the
cuffs open so as to
prevent a blockade of
the arteries.
Use cuff extensions. These are
available at
www.venenwalker.com
.