L060-0082-001 Rev. A
27 Radio Circle Drive
Mt. Kisco, NY 10549 USA
1-888-652-6520 • www.vistalab.com
5. Use the filter insertion tool to push the
filter into position. Press down until
the end of the tool makes contact with
the inside bottom of the nozzle.
6. Tap the nozzle on the counter to
remove any loose cellulose material.
7. Place rubber tubing back onto the
nozzle, and screw the nozzle onto the
pipette in a clockwise direction. Firmly
tighten and remove tubing from nozzle.
Special Note for 0.5–20µL Models:
To remove a nozzle(s) on 0.5–20µL
models, the rear panel of the ejector
sleeve needs to be removed.
Remove two screws on the panel, and
slide it away as shown. Remove springs
from posts, and remove affected nozzle(s)
with tubing provided.
After filter(s) has been replaced, reinstall
springs, rear panel and screws.
Nozzle Filter Replacement 10/5/04 3:16 PM Page 2