When to Replace:

Nozzle filters are made of a cellulose
material which prevents liquid from
entering the interior of the pipette. If
the filter becomes wet, the channel will
not aspirate or dispense until the filter 
is replaced.

Catalog No. 9060-4005

Package includes:

• 48 nozzle filters

• Tubing

• Filter insertion tool

• Instructions

Note: Same nozzle filter is used for all models

Multichannel Nozzle Filter
Replacement Instructions

Note:  Wear gloves

1. Push the ejector slide in until it clicks

to fully expose the pipette’s nozzles.

(For 0.5-20µL models, see note on reverse)

2. Remove the affected nozzle(s) by

placing the rubber tubing provided
onto nozzle, and unscrew it in a
counterclockwise direction.

3. Remove the filter inside the nozzle(s)

by inserting a straightened paperclip
into the small hole in the tip end of
the nozzle, and push the filter out
through the wide (threaded) opening.

4. Pick up a new nozzle filter with

tweezers and place it into the wide
(threaded) end of the nozzle.

NOTE: Make sure the narrow end is

inserted first 

Procedure: