Note:  Wear gloves

1. Push the ejector slide in until it clicks

to fully expose the pipette’s nozzles.

(For 0.5-20µL models, see note on reverse)

2. Remove the affected nozzle(s) by

placing the rubber tubing provided
onto nozzle, and unscrew it in a
counterclockwise direction.

3. Remove the filter inside the nozzle(s)

by inserting a straightened paperclip
into the small hole in the tip end of
the nozzle, and push the filter out
through the wide (threaded) opening.

4. Pick up a new nozzle filter with

tweezers and place it into the wide
(threaded) end of the nozzle.

NOTE: Make sure the narrow end is

inserted first 

5. Use the filter insertion tool to push

the filter into position.  Press down
until the end of the tool makes
contact with the inside bottom of the
nozzle.

Procedure:

Nozzle Filter Replacement Instructions

6. Tap the nozzle on the counter to

remove any loose cellulose material.

7. Place rubber tubing back onto the

nozzle, and screw the nozzle onto the
pipette in a clockwise direction.  Firmly
tighten and remove tubing from nozzle.

Special Note for 0.5–20µL Models:

To remove a nozzle(s) on 0.5–20µL
models, the rear panel of the ejector
sleeve needs to be removed.

Remove two screws on the panel, and
slide it away as shown.  Remove springs
from posts, and remove affected nozzle(s)
with tubing provided. 

After filter(s) has been replaced, reinstall
springs, rear panel and screws.

2–125µL 

0.5–20µL

5–250µL

25–850µL

Nozzle Reconditioning Kit

9060-1011

9060-2011

9060-3011

Contains 2 nozzles, 24 seals, 24 filters & tool

M

U L T I C H A N N E L

M

O D E L

Applicable Nozzle Reconditioning Kits

9060-1011

Catalog Number