Multichannel Nozzle
Reconditioning Kit
Note: Wear gloves
1. Push the ejector slide in until it clicks
to fully expose the pipette’s nozzles.
2. Remove an existing seal by slipping
the end of a very small screwdriver or
tweezers under the edge of seal,
taking care not to score or scratch
nozzle surface. Stretch it up and over
the end of the nozzle.
3. Place a new seal on the countertop,
with the rib line closest to counter
surface.
4. Pick up the pipette and align the
nozzle with the new seal. Press down
to ease the seal over the end of the
nozzle and into place.
Procedure:
Nozzle Seal Replacement Instructions
L060-0100-001 Rev. B
2 Geneva Road
Brewster, NY 10509 USA
1-888-652-6520 • www.vistalab.com
The Nozzle Reconditioning Kit contains
complete nozzles and components that
can be replaced as needed.
Applicable Catalog Numbers:
9060-1011
9060-2011
9060-3011
Package includes:
• 2 complete nozzle assemblies
• 24 replacement filters
• 24 replacement seals
• Filter insertion tool
• Tubing
• Instructions
Note: Same nozzle filter is used for all models
When to replace Nozzles:
If a nozzle is leaking or does not aspirate, a complete nozzle can be replaced.
When to replace Nozzle Filters:
Nozzle filters are made of a cellulose material which prevents liquid from
entering the interior of the pipette. If the filter becomes wet, the channel will
not aspirate or dispense until the filter is replaced.
When to replace Nozzle Seals:
Nozzle seals may need to be replaced when a channel leaks during pipetting,
or the seal is visibly worn or torn.
1. Push the ejector slide in until it clicks to
fully expose the pipette’s nozzles.
(For 0.5-20µL models, see note on reverse)
2. Remove the affected nozzle(s) by placing
the rubber tubing provided onto nozzle,
and unscrew it in a counterclockwise
direction.
3. Place rubber tubing back onto a new or
reconditioned nozzle, and screw the
nozzle onto the pipette in a clockwise
direction. Firmly tighten and remove
tubing from nozzle.
Nozzle Replacement Instructions
Note: Wear gloves