VistaLab Ovation 9060-1011 Скачать руководство пользователя страница 1

Multichannel Nozzle
Reconditioning Kit

Note:  Wear gloves

1. Push the ejector slide in until it clicks

to fully expose the pipette’s nozzles.

2. Remove an existing seal by slipping

the end of a very small screwdriver or
tweezers under the edge of seal,
taking care not to score or scratch
nozzle surface. Stretch it up and over
the end of the nozzle.

3. Place a new seal on the countertop,

with the rib line closest to counter
surface.

4. Pick up the pipette and align the

nozzle with the new seal.  Press down
to ease the seal over the end of the
nozzle and into place.

Procedure:

Nozzle Seal Replacement Instructions

L060-0100-001 Rev. B

2 Geneva Road
Brewster, NY 10509  USA
1-888-652-6520  •  www.vistalab.com

The Nozzle Reconditioning Kit contains
complete nozzles and components that
can be replaced as needed.  

Applicable Catalog Numbers:

9060-1011
9060-2011
9060-3011

Package includes:

• 2 complete nozzle assemblies

• 24 replacement filters

• 24 replacement seals

• Filter insertion tool

• Tubing

• Instructions

Note: Same nozzle filter is used for all models

When to replace Nozzles:

If a nozzle is leaking or does not aspirate, a complete nozzle can be replaced.

When to replace Nozzle Filters:

Nozzle filters are made of a cellulose material which prevents liquid from
entering the interior of the pipette. If the filter becomes wet, the channel will
not aspirate or dispense until the filter is replaced.

When to replace Nozzle Seals:

Nozzle seals may need to be replaced when a channel leaks during pipetting,
or the seal is visibly worn or torn.

1. Push the ejector slide in until it clicks to

fully expose the pipette’s nozzles.

(For 0.5-20µL models, see note on reverse)

2. Remove the affected nozzle(s) by placing

the rubber tubing provided onto nozzle,
and unscrew it in a counterclockwise
direction.

3. Place rubber tubing back onto a new or

reconditioned nozzle, and screw the
nozzle onto the pipette in a clockwise
direction.  Firmly tighten and remove
tubing from nozzle.

Nozzle Replacement Instructions

Note:  Wear gloves