11.5.
Connecting the BaseSensor
The BaseSpion is
connected to the
LabSensor with a RJ45
cable.
A 7.5-meter RJ45 Cat5
shielded cable is supplied
with the BaseSpion, but
any shielded RJ45 cable
can be used.
Warning: Do not connect the Sensor to the C-plane motor connector, this could
damage the Sensor.
11.6.
Connecting the C-plane Goniometer
The C-plane goniometer is connected to the BaseSpion base through a RJ45 cable.
The BaseSpion will automatically detect the C-plane goniometer.
Warning: Do not connect the C-plane motor to the Sensor connector. This could
damage the BaseSpion.
11.7.
Connecting Light Source Power
The BaseSpion has a built-in power analyzer and power switch. The power switch is
used when running in ambient light correction mode. So the lamp can be switched
off before a measurement, so that the values of the ambient light can be obtained
and subsequently subtracted from final measurements.