Adjusting the Swing out Condenser (Fig. 32) 

The  centre  of  the  condenser  and  the  light  axes  of  the  objective  are 
coaxial.  This  has  been  adjusted  before  leaving  factory,  so  the  user 
doesn’t need to adjust them. 

The  condenser  has  been  adjusted  to  its  highest  setting.  It  does  not 
require the user to manipulate this feature. 

To  move  the  condenser,  please  use  the  ‘condenser  focus  control’.   
When using the high magnification objective, make sure the condenser 
is raised; similarly, if you’re using a low magnification objective make 
sure the condenser is lowered. 

1.

 

Using the Swing out Condenser 

2.

 

When  using  the  low  magnification  objective,  make  sure  the 
condenser  is  away  from  the  light  path.  While  using  the  high 
magnification objective, make sure this is in the light path. 

 

Adjusting the Aperture Diaphragm 

The  aperture  diaphragm  is  designed  for  the  adjustment  of  the 
numerical  aperture,  not  for  the  brightness.  Generally,  reducing  the 
diaphragm  opening  to  70  -  80%  of  the  N.A.  value  of  the  respective 
objective  will  provide  an  image  of  acceptable  quality.  If  you  want  to 
observe  the  image  of  the  aperture  diaphragm,  remove  one  eyepiece 
and look through the tube. You will see a dark circle encroaching on 
the bottom of the tube. 

 

3-15 

Adjusting the Field Diaphragm (Fig. 33) 

The control for the field diaphragm is a ring used for adjusting the area 
of  field  diaphragm.  When  using,  turn  the  ring  to  reduce  the  field 
diaphragm,  look  into  the  field,  if  the  diaphragm  image  is  faint, 
complete  the  follow  steps:  turn  the  condenser  focus  knob,  shift  the 
condenser holder to the position where the observed image in the field 
of view is sharp; then open the field diaphragm, let the image fill the 
field  of  view,  then  reduce  the  mixed  light,  which  should  improve  the 
quality of the image. 

Fig.32 

Fig.33 

Содержание TIM5

Страница 1: ...TIM5 User Guide Material Inspection Microscope...

Страница 2: ...1 Contents USER NOTICE 2 1 Components 3 2 Assembly 4 3 Operation 9 4 Observation Methods 16 5 Troubleshooting 18...

Страница 3: ...lamp house completely cool before removal vi Use the factory supplied power cord II Maintenance i Do not disassemble any parts of the microscope as it will affect function or reduce the performance ii...

Страница 4: ...nsertion slot Aperture diaphragm lever Field diaphragm lever Main switch Brightness adjustment knob Right fine focus knob Mechanical stage X Y stage movement Lamp housing Trinocular tube Filter insert...

Страница 5: ...allation steps Before installing be sure every component remains clean Note The supplied Allen key is required for changing components Trinocular tube Trinocular head 10x wide field eyepiece Lamp hous...

Страница 6: ...so you can install the mechanical stage support device easily Hold on to the mechanical stage support device Fig 2 Place it from the top of the guide board Fig 1 until the stage reaches the lowest po...

Страница 7: ...the objective 1 Adjust the coarse focus knob until the support device of the mechanical stage reaches its lower limit position 2 Place the lowest magnification objective onto the nosepiece from the le...

Страница 8: ...Fig 10 into the illumination kit Fig 4 turn it to the correct position and then screw down bolt to fix it into position 2 Insert the power plug into the jack socket and screw in the bolt Remember to...

Страница 9: ...bulb during or directly after use Be aware that during use and directly after the equipment has been turned off the lamp the lamp housing and nearby parts will be very hot and could cause serious bur...

Страница 10: ...the brightness turn knob clockwise to increase the voltage therefore increasing the illumination and turn anti clockwise to turn down the voltage and dim the light Using the lamp at low voltage will...

Страница 11: ...e and do not allow the objective to touch the specimen This could cause damage to the lens 3 5 Adjusting the Interpupillary Distance The interpupillary distance range between 48mm 75mm Make sure when...

Страница 12: ...e specimen whilst viewing at 10x then the user can adjust the fine control to focus more clearly on the object At this point the user can change the objective to a higher magnification without the ris...

Страница 13: ...to the mirror selector lever in Fig 25 You can reduce the glare with the ND filter when you switch from darkfield DF to brightfield BF Releasing interlocking The ND filter control has been locked at t...

Страница 14: ...so that the field iris image becomes concentric within the field of view 5 While pushing in the FS control Fig 27 open the field iris diaphragm until the field iris image inscribes the field of view...

Страница 15: ...pened by pushing in the AS knob With some specimens an image with high contrast and little flare may sometimes be obtained when the aperture is slightly closed It is therefore recommended to also try...

Страница 16: ...rture Diaphragm The aperture diaphragm is designed for the adjustment of the numerical aperture not for the brightness Generally reducing the diaphragm opening to 70 80 of the N A value of the respect...

Страница 17: ...ain switch to I ON 2 1 Disengage the analyser polariser filter etc from the light path 2 2 Check interlocking of the ND filter 3 Select the light path 4 Place the specimen on the stage 5 Engage the 10...

Страница 18: ...the field iris diaphragm so that the image fills the field of view 5 You may need to close the aperture iris diaphragm slightly to enhance the contrast 4 3 Transmitted Light Observation 1 Installing t...

Страница 19: ...ion Set it into the click position Field iris diaphragm is not centred Centre the field iris diaphragm correctly and make sure it is sufficiently open ND filter is not in a click position Set it into...

Страница 20: ...rely Field of view is not the same in both eyes Interpupillary distance is incorrect Adjust the interpupillary distance The eyes tire easily Incorrect adjustment to the dioptre Adjust dioptre correctl...

Страница 21: ...20 LIT5072 R1 0 0814...

Отзывы: