88
89
Visiomed - KD-795 - Manuel d’utilisation - 022019
Visiomed - KD-795 - Manuel d’utilisation - 022019
ELECTROMAGNETIC COMPATIBILITY INFORMATION
Table 1 - Emission
Phenomenon
Compliance
Electromagnetic environment
RF emissions
CISPR 11
Group 1, Class B
Home healthcare environment
Harmonic
distortion
IEC 61000-3-2
Class A
Voltage fluctuations
and flicker
IEC 61000-3-3
Compliance
Table 2 - Enclosure Port
Phenomenon
Basic EMC standard
Immunity test levels
Home Healthcare Environment
Electrostatic
Discharge
IEC 61000-4-2
±8 kV contact
±2kV, ±4kV, ±8kV, ±15kV air
Radiated RF EM field
IEC 61000-4-3
10V/m
80MHz-2.7GHz
80% AM at 1kHz
Proximity fields from
RF wireless communi-
cations equipment
IEC 61000-4-3
Refer to table 3
Rated power
frequency magnetic
fields
IEC 61000-4-8
30A/m
50Hz or 60Hz
Table 3 – Proximity fields from RF wireless communications equipment
Test frequency
(MHz)
Band (MHz)
Immunity test levels
Professional healthcare
facility environment
385
380-390
Pulse modulation 18Hz, 27V/m
450
430-470
FM, ±5kHz deviation, 1kHz sine,
28V/m
710
704-787
Pulse modulation 217Hz, 9V/m
745
780
810
800-960
Pulse modulation 18Hz, 28V/m
870
930
1720
1700-1990
Pulse modulation 217Hz, 28V/m
1845
1970
2450
2400-2570
Pulse modulation 217Hz, 28V/m