SAFETY PRECAUTIONS 

 

All safety messages are identified by the following the words "

WARNING

" and 

"CAUTION". These words mean the following: 

 

WARNING

 

Important information to alert you to a 
situation that might cause serious injury and 
damage to your property if instructions are 
not followed. 

CAUTION 

Important information that tells how to prevent 
damage to equipment. 

 

When using the scale/balance, the following safety precautions should always be 
followed. 
 

△  

WARNING 

Use only the AC adaptor with the scale/balance. Other adaptor may cause 

damage. (AC adaptor is optional) 

 

△  

CAUTION 

 

Avoid installing the scale/balance in direct sunlight, this may cause discoloration or 
malfunctions. 

 

Do not mix the type of batteries. Replace all other batteries at the same time. 

 

If the scale/balance is not to be used for a long period of time, remove all batteries 

from the battery compartment to avoid leakage which may cause damage to the 

instrument. 

 

Avoid overloading that could cause damage to the scale/balance. 

 

Do not drop water in the scale/balance that is not water-resistant. It causes damage, 

if inside of the scale/balance is wet. 

 

Matter charged static electricity could influence the weighing. Discharge the static 

electricity. As example, the method is to use the electrification prevention spray, to 

spray it to both side of weighing platform. 

 

7.

 Place the items that you want to count onto the tray and the total number of items 

will show on the display. 

8.

 Press the 

[UNIT]

 key to exit the counting function and return to normal weighing 

or you can press ON/OFF to turn the scale off to exit. 
NOTE: the weight of unit sample > 10e. 

CALIBRATION 

When to calibrate 

Calibration may be required when it is initially installed, if the scale/balance is 
moved to a substantial distance.This is necessary because the weight of a mass in 
one location is not necessarily the same in another location. Also, with time and use, 
mechanical deviations may occur. 

How to calibrate 

1. 

Before entering calibration mode, press 

[ON/OFF]

 to turn the power on for 5 

minutes.

 

2. select the calibration unit 
 

Press 

[ON/OFF]

 to turn the power off. Press and hold [

UNIT

] first ,then 

press[

ON/OFF

],the display will show “   ”,the 

sign indicate the  calibration 

unit .Press 

[ZERO] 

key, the 

  sign indicate the different calibration unit (eg: g

oz or lb) .Now the calibration unit is selected.

 

3. Enter  calibration  mode 

Press 

[ON/OFF]

 to turn the power off. Press 

[UNIT]

 first, then Press

 [ON/OFF]

while keeping pressing 

[UNIT]

, the display will show “CAL”then press the

[UNIT] 

again

,

the display show”888888” and then display all segments, finally show the 

A/D value . 

4. Calibration  by  the  weight 

Wait for the stable indicator to be displayed, Press 

[UNIT],

 waiting for 3 seconds, 

the display will show 

"00SAVE"

,then the display show the calibration weigh ,Place 

the    calibration weight on platform,    after the stable indicator is displayed, then 
press 

[UNIT]

, waiting for 3 seconds, the display will show 

"00SAVE"

and return to 

A/D value, now calibration is finished.

 

5. 

Press 

[ON/OFF]

, turn off the power, press 

[ON/OFF]

 turn on the power again, 

check some weight readings , if not, repeat steps 3-5.

 

6. Return to weighing mode   

Press 

[ON/OFF]

 to turn the power off. Press 

[ON/OFF]

 again to turn on the 

power and the scale/balance return to the weighing mode. 

 

 

 

 

 

 

6

 

CAL

Содержание VB-302

Страница 1: ...ing Auto backlight selectable Low batter indication g oz lb dwt tl H tl T ct ozt pcs Large LCD 6digits 15 mm high with counting function Stability indication http www balances com PRECISION BALANCE MODEL VB 302 www balances com is your deep discount distributor INSTRUCTION MANUAL USA ASIA Virtual Measurements control Inc Virtual Measurements control Asia Pte Ltd 3196 Coffey Lane Suite 604 24 Hillv...

Страница 2: ...ng or you can press ON OFF to turn the scale off to exit NOTE the weight of unit sample 10e CALIBRATION When to calibrate Calibration may be required when it is initially installed if the scale balance is moved to a substantial distance This is necessary because the weight of a mass in one location is not necessarily the same in another location Also with time and use mechanical deviations may occ...

Страница 3: ...he display to 0 3 Place a given number of samples of an item on the pan the Sample Size should be either 10 20 50 or 100 pieces The weight of these samples will show on the display 4 Press the UNIT key several times to put the scale in PCS mode 5 Select the sample size the same as you chose above by pressing the PCS key press it as many times as necessary to put it in the correct sample size the s...

Страница 4: ... please allow the balance to warm up for 10 minutes after first turning to power on so that the balance will function Properly and accurately Error messages EEEE Overload Low battery Calibration may be required before weighing Read CALIBRATION first and if necessary calibrate your scale balance for accurate weighing Weighing procedures 1 Press ON OFF to turn on the scale balance When power is turn...

Отзывы: