22
Rev. 1 20.12.2019
EN
second screen: CALIBRATION DATE and TIME (if a GLP instrument is used) and beakers indicating
the buffers used;
third screen: OFFSET value of the electrode expressed in mV;
fourth and possibly fifth screen: Slope% in the measuring range (one Slope% only if two calibration
points are performed, two Slope% if three points are performed).
From the moment the DHS electrode is recognized, the device use the calibration stored on the probe.
The icon on the display
indicates that the connection was successful.
If the calibration is satisfactory (see the calibration data in menu P.1.6) the electrode is ready to start
the measurements. Otherwise recalibrate the electrode; the data will be updated automatically.
The DHS electrode calibrated with a pH 7 Vio or PC 7 Vio device is ready to be used on any pH meter
enabled for DHS recognition and vice versa.
When the electrode is disconnected, a message on the display informs the user of the deactivation of
the sensor; the device use its previous calibration stored in the device, no data is lost!
The DHS electrode does not require batteries and if it is used on pH meters that are not enabled to
recognize the chip, it works as a normal analog electrode.
Errors during calibration
NOT STABLE: The button
has been pressed with still unstable signal. Wait for the icon
to
appear to confirm the point.
WRONG BUFFER: The buffer you are using is polluted or not part of the recognized families.
SLOPE OUT OF RANGE: The slope of the sensor calibration line is out of the acceptable range 80 - 120%.
CALIBRATION TOO LONG: the calibration has exceeded the time limit, only the points calibrated up to
that moment will be kept.
mV Parameter
pH 7 Vio; PC 7 Vio
In measurement mode press
and move to the mV parameter indicated by the icon
The display shows the measurement in mV of the pH sensor.
Scroll on the display of four red bands
means that the measurement is not yet stable.
Consider the measurement truthful only when the stability icon appears
.
Note: This measurement is recommended to evaluate the sensor efficiency
ORP Parameter (Redox Potential)
pH 7 Vio; PC 7 Vio
ORP sensors can be used on this series of devices to measure the Oxide-Reduction Potential.
Connect the Redox electrode to the BNC type connector marked in green; instead, if necessary, connect the
temperature probe to the RCA / CINCH Temp connector always marked with a green background.
It is possible to calibrate the sensor offset by going for automatic calibration to a predefined point. The
instrument automatically recognizes the Redox solution 475 mV / 25 ° C; contact the local distributor to
proceed with the purchase.
The instrument can correct the sensor offset by + 75 mV.
Содержание 7 Vio
Страница 1: ...1 Rev 1 20 12 2019 EN 7 Vio INSTRUCTIONS MANUAL pH Cond PC ...
Страница 2: ...2 Rev 1 20 12 2019 EN ...