Balance mechanism digital display calibration
The digital display
indicates setting of the balance mechanism on a scale of 00 (minimum set-
ting) to HI (maximum setting). In the unlikely event of this system requiring calibration, proceed as
follows:
Level the platform and apply centre lock
.
fully clockwise to its maximum stop.
Press and hold the switch
for approximately eight seconds, until the digital display
shows CA. Release the switch
. The display shows HI.
With the balance knob
still turned fully clockwise, press and release the switch
display will now show LO.
fully counter-clockwise to its minimum stop.
Press and release switch
.
If the calibration is successful, the display will now show 00.
If unsuccessful, Er will be displayed. Pressing the switch
again, or waiting for 15 sec-
onds, will allow the system to revert to its previous settings. Calibration may now be carried
out again.
After calibration, rebalance the head (see Balancing the head).
NOTE:
If more than 15 seconds is allowed to elapse between steps, the system will
shut down and revert to its previous settings.