vinmetrica SC-100A Скачать руководство пользователя страница 4

SC-100A manual 

Version 3.1c 

Instrument Operation 

 

Turn on the instrument by pressing the POWER

 

button briefly (Note: depressing the POWER

 

 button 

longer than two seconds at start-up will cause the instrument to enter 

Test Mode;

 see Appendix A). The 

instrument will go through a power-up sequence. After a few seconds the display will show a value less 
than 50 (usually 0.0) and the green (Proceed) LED should be on.  

 

Procedures

  

 

Measuring Free SO

2

 by Titration:  

 

 
1.

 

Turn on the instrument and connect the electrode. The display should show a value less than 20, 
usually 0.0. 

 

2.

 

Fill the syringe by drawing up the SO

2

 Titrant (the bottle with the blue dot) 

(Figure 4). Expel bubbles and set the plunger on the syringe to a readable point, 
preferably the 5.0 mL point. 

This is your starting volume

. Make sure the outside 

of the syringe is dry to minimize inaccuracies. [Note: the 5.0 mL setting allows 
determination of up to 100 ppm SO

2  

in a standard 25 mL wine sample.] If using a 

burette, us the syringe to dispense the SO

Titrant into the top of the burette.

 

 

3.

 

Using the 25 mL pipette provided in the kit, place 25 mL of wine or must in the 
titration beaker (Figure 5). Draw sample up to the 0 mL mark, then dispense the 
sample into your titration beaker by letting the tip of the pipette touch the side of 
the vessel while the sample drains. 

NEVER pipette any reagents by mouth! 

Also make sure the pipette you are using is completely clean before 
submerging into your wine sample.

 –Do not use sulfite to sanitize it! Use 25% 

ethanol or other sterilent.  

 
 
 
 
 
 
 
 
 
4.

 

Using the transfer pipettes, add about 2 mL Acid Solution and 2 mL Reactant solution to the 
titration beaker. It is not necessary to be extremely accurate in this step; with these pipettes, 2 mL 
is indicated on the neck of the pipette. Please make sure to label the pipettes to eliminate cross 
contamination, this will prolong the shelf life of the reagents e.g. mark the pipette for the Acid 

Figure 4. 

Withdraw 

the titrant from its 

bottle using a clean 5 

mL syringe.  

Figure 5.

 Dispense 25 

mL of your wine into the 
titration beaker using the 

25 mL sampling pipette. 

Make sure this is clean 

before putting the pipette 

into your wine container! 

Содержание SC-100A

Страница 1: ...ge 5 Two 3 mL polyethylene transfer pipettes 6 One 25 mL serological pipette 7 One 100 mL polypropylene beaker Things you will need 1 Two standard AA batteries alkaline type 2 Distilled water which us...

Страница 2: ...63 79 99 125 pH 3 0 3 1 3 2 3 3 3 4 3 5 3 6 3 7 3 8 3 9 4 0 We recommend using a sulfite calculator for determining how much sulfite to add to your wines after taking a sulfite measurement with the SC...

Страница 3: ...100A shuts off after 30 minutes If this happens unexpectedly just press the POWER button to resume from where you were 4 The SO2 electrode is attached via the BNC connector protruding from the top To...

Страница 4: ...to dispense the SO2 Titrant into the top of the burette 3 Using the 25 mL pipette provided in the kit place 25 mL of wine or must in the titration beaker Figure 5 Draw sample up to the 0 mL mark then...

Страница 5: ...lectrode Holder adjust it to a similar level You may add up to 20 mL water if needed 8 Verify that the current is less than 50 and the green PROCEED LED is lit If the current is greater than this and...

Страница 6: ...gent and 2 mL of the Reactant solution to the vessel Remember that if you are using the transfer pipettes in the kit 2 mL is marked on the side of the neck of the transfer pipette so dispense four of...

Страница 7: ...METRICA S MAXIMUM LIABILITY FOR ALL DIRECT DAMAGES INCLUDING WITHOUT LIMITATION CONTRACT DAMAGES AND DAMAGES FOR INJURIES TO PERSONS OR PROPERTY WHETHER ARISING FROM VINMETRICA S BREACH OF THESE TERMS...

Страница 8: ...st Mode hold the POWER button down 5 10 seconds until the instrument shuts off If the device does not shut off after 10 seconds of holding down the button move to the next section by releasing then pr...

Страница 9: ...2 pH CAL values None only firmware v3 0 6 and later NA Displays current CAL values for pH 7 and 4 in mV Pressing ENTER cycles between these If a pH CAL reset has been done displays CAL value for pH 3...

Страница 10: ...r this type of wine molecular SO2 pH is your value mg L Redo the calculation using this value for desired free SO2 level if required Enter this value in the Desired level of free SO2 box Press Calcula...

Страница 11: ...ulated I m getting strange results in SO2 mode how do I know if my instrument is working correctly For SO2 measurements with the SC 100A there are several quick tests you can do to make sure the instr...

Страница 12: ...connector to its outer metal sheath The device should indicate STOP with its red LED and buzzer or beeper and the current should go to 1999 If this does not happen there may be a problem with the ins...

Отзывы: