IM 04L75B01-01EN

LOG Input Function

Foreword

This section explains the 

LOG input function

For a description of other standard functions, see the 

DX2000 User’s Manual (IM04L42B01-01E) and the 

Communication Interface User’s Manual (IM04L41B01-

17E).

*  The accompanying software application 

DAQSTANDARD for DXAdvanced is the special 

version for the DX3000 with these special functions.

  The setting function for these special functions and 

the saving setting data by ASCII or Excel format are 

available on this special software.

LOG Input Specifications

■  Converting the LOG Input (Relationship between 

the Input Voltage and Digital Display)

Trend display

(logarithmic scale)

:  Displays the input voltage on a logarithmic 

scale.

Digital display

:  Displays the result obtained by converting 

the input voltage using the equation below.

The relationship between the input voltage of special LOG 

range and digital display value is shown below.

VL

VU

Y

1.0E

SL

1.0E

SU

Input X

Input voltage
Lower limit of display span
Scale lower limit 

Upper limit of display span
Scale upper limit 

: X
: VL
: SL

: VU
: SU

Digital display value Y=10

(SU-SL)*(X-VL)/(VU-VL)+SL

■  LOG Input Mode

•  This mode allows you to measure the DC voltage input 

(DCV: 20 mV, 60 mV, 200 mV, 2 V, 6 V, 20 V, and 50 

V), convert to logarithmic scale, and display the data.

•  Selectable span range: Same range as the DC voltage 

input

  Span Lower must be less than Span Upper.

•  Selectable scale range: 1.0E-15 to 1.0E+15 (15 

decades maximum)

  Scale Lower must be less than Scale Upper.

•  Select two or three digits for the mantissa.

•  You can set the unit.

 

Setting the LOG Input

1.

  Press 

MENU

 (switch to the setting mode).

2.

  Select 

Meas channel

 > 

Range, Alarm

.

3.

  Select the channels in 

First-CH

 and 

Last-CH

.

4.

 Set 

Mode

 to Log.

5.

  Set the measurement range, the span, and the 

exponents of the scale lower and upper limits.

LOG Input Setup Screen Example

Setup Item

Selectable Items

(Selectable Range)

Description

Range

Same as the DCV

range

Sets the range.

Span Lower Varies depending on 

the input type

Sets the span lower limit.

Span Upper Varies depending on 

the input type

Sets the span upper limit.

* Span Lower must be less 

than Span Upper.

Scale Lower 1.0E-15 to 1.0E+15 Sets the scale lower limit.

1.00E-15 to 1.00E+15 when 

the number of displayed 

mantissa digits is 3.

Scale Upper 1.0E-15 to 1.0E+15 Sets the scale upper limit.

* 15 decades maximum.

*  Scale Lower must be less 

than Scale Upper.

1.00E-15 to 1.00E+15 when 

the number of displayed 

mantissa digits is 3.

Unit

Up to 6 characters

Up to 6 characters

Table 1 Log Input Mode Settings

Содержание Daqstation DX3004

Страница 1: ...DX3030 Daqstation DX3000 User s Manual Contents Introduction 1 Specifications Different from the DX2000 3 Checking the Package Contents 3 External Dimensions and Panel Cut Dimensions 5 Names of Parts...

Страница 2: ...gistration Thank you for purchasing YOKOGAWA products We invite you to register your products in order to receive the most up to date product information To register visit the following URL http www y...

Страница 3: ...4EN DXA120 Installing DAQSTANDARD IM 04L41B01 66EN Precaution on the use of the DX3000 IM 04L75B01 91EN Genaral Specifications Title General specifications No Paperless Recorder DX3000 GS 04L75B01 01E...

Страница 4: ...ES INCLUDING BUT NOT LIMITED TO PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES LOSS OF USE DATA OR PROFITS OR BUSINESS INTERRUPTION HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY WHETHER IN CONTRACT STRIC...

Страница 5: ...Dimension and Weight See External Dimensions and Panel Cut Dimensions Vibration 5 f 8 4 Hz amplitude 3 5 mm peak 8 4 f 160 Hz acceleration 9 8 m s2 The accompanying software application DAQSTANDARD fo...

Страница 6: ...RD IM 4L41B01 66EN 1 A4 size 4 CF card B8706NQ 1 128 MB The size and model may change Optional Accessories Sold separately The following optional accessories are available for purchase separately If y...

Страница 7: ...43 5 5 65 196 7 72 71 5 2 81 Ddimensions 480 4 min 18 91 195 3 2 0 7 69 435 9 2 0 17 16 227 1 min 8 94 7 5 0 3 10 5 0 41 Allowable panel thickness Unit mm approx inch Unless otherwise specified tolera...

Страница 8: ...ator Key panel Key panel Power switch Rear Panel USB port USB1 option Serial interface port C2 option A RS 232 interface connector Input terminals screw terminals Connect input signal wires from the m...

Страница 9: ...measure the DC voltage input DCV 20 mV 60 mV 200 mV 2 V 6 V 20 V and 50 V convert to logarithmic scale and display the data Selectable span range Same range as the DC voltage input Span Lower must be...

Страница 10: ...span value of the range to be 20000 The full span value of the range is the maximum value that can be set in the range It is 6 V in the 6 V range For example the A D normalized values in the 6 V range...

Страница 11: ...t voltage X Lower limit of display span VL Upper limit of display span VU Scale lower limit SL Scale upper limit SU Digital display value Y 1 Converting equation for the input voltage setting Y 10 SU...

Страница 12: ...ased in this condition the remote control input terminal will be open even though the memory is in a start condition Action Note Memory start stop Does not operate while the display is frozen Operates...

Страница 13: ...key Setup Screen Example for the Memory and Trend Clear Function Communication Commands The same command is used to control the display freeze function and memory and trend clear function QB p1 termin...

Страница 14: ...rent value mark setting is in AUX setting of the basic setting mode See Table 6 Fig 1 Current Value Mark Channel Type Channel Display of the Current Value Mark Three Digits Standard Two Digits Special...

Страница 15: ...p Function for the Number of Displayed Mantissa Digits The setting for the number of displayed mantissa digits is added to the Aux area of the Basic Setting page tab For details on the settings see th...

Страница 16: ...Blank...

Страница 17: ...Floor 129090 Moscow RUSSIA Phone 7 495 737 7868 Fax 7 495 737 7869 YOKOGAWA CHINA CO LTD 3F Tower D Cartelo Crocodile Building No 568 West Tianshan Road Shanghai 200335 CHINA Phone 86 21 62396262 Fax...

Отзывы: