VIDITECH Vibrodiag Скачать руководство пользователя страница 18

VIBRODIAG

 

 

www.viditech.eu 

sales

@viditech.cz

 

 

17

 

Fig. 16  

c) Graph options. 

 

 

 

d) Coast down display. 

Saved  records  can  be  displayed  using  Saved  data  (fig.  16a).  After  this  selection,  two 
columns are added to the saved items (ranged by sensing time – from the newest). The 
first column is for data sensed during run up and the second column during machine coast 
down. Tapping on the required box shows channel options and graph type options. Graph 
control is analogical to the previous cases, see fig. 16c. It is possible to add new records 
into the graph and to compare them among each other. 

 

7.4

 

Balancing 

For correct and secure instrument  balancing, minimally the following  instructions must 
be followed: 

a)

 

Place on the machine only weights specified by the manufacturer, into spots that 
are  explicitly  designated  for  this  purpose  –  balancing  axis  (cutting,  thread  holes, 
places for welding and other) 

b)

 

During  the  weight  placement  selection,  safety  and  health  of  the  operating 
personnel  must  be  taken  into  account,  against  weight  detachment  due  to 
centrifugal force. 

c)

 

A larger weight, than the one already placed on the machine, should not be used. 
Rather choose a larger number of lighter weights, and repeat the measurement. 

d)

 

Personnel without appropriate practice should not conduct these measurements! 

7.4.1

 

Balancing in single plane 

Main menu 

 Measuring 

 Balancing 

 Single plane  

This  option  is  used  for  balancing  thin  rotating  wheels  on  short  shafts  of  rotating 
machines.  For  correct  function,  the  speed  sensor  must  be  connected.  Further,  it  is 
necessary  to  label  the  zero  degree  on  the  circular  (angular)  scale  of  the  rotating  wheel 
with a reflective band. From this reference position (0°), the following runs are calculated. 
After executing the process, screen with selection of previously saved machine coefficient 
appears. You can choose the coefficients from database (balancing without the test run) 
or do the measuring without choosing the coefficient from database (balancing with the 
test  run).  After  initiation  the  wizard  is  displayed,  it  asks  to  enter  the  machine  rotation 
direction and then to enter the position of the accelerometer towards the  speed sensor. 
The position is then noted in the balancing results. 

The  wizard  in  this  case  is  informative  and  used  for  better  orientation.  The  balancing 
result using the test weight does not depend on the wizard setup. It is possible to skip the 
wizard using Skip wizard and further the balancing continues as follows: 

First  run:  Tapping  on  the  forward  arrow  button,  vibration  analysis  of  rotating 

wheel  is  initiated.  For  the  analysis  to  be  performed,  the  flywheel  must  be  steady-state 

Содержание Vibrodiag

Страница 1: ...www viditech eu rev 1 7 sales viditech cz www ViDiTech eu Vibrodiag FW 0 04 41 and higher...

Страница 2: ...MENU 8 5 1 INSTRUMENT SETUP 9 5 2 SENSOR SETUP 10 5 3 LIMITS SETUP 10 6 DATABASE ROUTES 11 6 1 MEASURING PLACES 11 6 2 ROUTES 11 7 MEASURING 13 7 1 ROUTE MEASURING 13 7 2 ANALYSIS 14 7 2 1 Velocity an...

Страница 3: ...to 60g Range of measured velocities up to 200 m s 1 using provided sensors RMS vibration measurement according to ISO 2373 Balancing wizard Threshold recorder Periodical recorder ICP acceleration sens...

Страница 4: ...ch TFT Display diagonal 3 4 Memory card type SD PCconnection USB 2 0 2 2 Accuracy Total vibrations at80 Hz 2 1 digit within range 1 5 000 Hz 10 1digit at limit frequencies 10 20 1digit Frequencycharac...

Страница 5: ...nsor sensitivity 100 mV g 20 Range of measured accelerations 60 g Protection IP67 Chassis material stainless steel Dimensions 25x66 mm diameter height Weight 125 6 g Connector M12 Attachment outer thr...

Страница 6: ...channel A 5 Socket for ICP sensor channel B 6 Socket for revolution sensor 7 Socket for supply connector 8 Socket for USB connector 9 SD card slot 10 Power supply adapter 15 V 1 6 A 11 ICP sensor cabl...

Страница 7: ...ts the instrument stabilizes the filters for only 2 s 3 1 2 ICP sensors ICP sensors are connected with the unit using shielded cables with colour distinguished connectors for each channel The only and...

Страница 8: ...m Analysis Balancing Recorder VibroGuide Order 1 Hz 1 kHz velocity 500Hz 5kHz acceleration Custom Saved data Run up coast down Run up Coast down Saved data Single plane Dual plane N planes Saved data...

Страница 9: ...equently it is possible to save individual measurements which point to the selected places routes Item Measuring contains all options of measuring allowed by the purchased licences see Setup i VibroGu...

Страница 10: ...erm instrument non use there is a danger of set date and time loss After power resupply in form of charged battery or connected charger random date and time are set and is therefore necessary to updat...

Страница 11: ...imits setup Main menu Setup Limits Limits are used to advise of detected increased values of vibration velocity on the diagnosed object These increased vibrations refer to the current machine state Li...

Страница 12: ...by place definition of Machine using icon and place parameters setup using icon with switching screens left right see fig 10 Here it is possible to set parameters of channel A and B and predefined op...

Страница 13: ...sales viditech cz 12 Fig 9 a Factory definition b Machine definition c Machine information Fig 10 a Place definition b Place setup c Place setup Fig 11 a Route definition b Adding place to route c Di...

Страница 14: ...performed by tapping on the box next to the relevant channel Bar graphs change colour according to the level of signal in reference to the set limits L1 L2 If RMS velocity exceeds threshold L1 bar gra...

Страница 15: ...requency and amplitude see fig 14b After specifying the relevant value in the table row the setting will change back to graph and mark the selected point using a cursor Values in the table and graph a...

Страница 16: ...sales viditech cz 15 Fig 13 a Analysis selection b Channel selection c Running analysis Fig 14 a Analysis display b Table of values c Optional analysis Fig 15 a Saved analyses b Saved analysis c Addi...

Страница 17: ...ther with the current vibration velocity values are displayed in the bottom part of the screen After revolutions set in final state run up to maximum coast down to minimum it is necessary to end the r...

Страница 18: ...Personnel without appropriate practice should not conduct these measurements 7 4 1 Balancing in single plane Main menu Measuring Balancing Single plane This option is used for balancing thin rotating...

Страница 19: ...ancing weights it is necessary to remove the test weight A control may be done by performing a new balancing It is possible to return by one step back leave and save result in the balancing process Fi...

Страница 20: ...s sense it is the revolution frequency Test Run Sizing counterweight must be performed with a running test weight Test weights are attached to balancing wheels the weight and the position must be ente...

Страница 21: ...er the position of the accelerometer towards the speed sensor The position is then noted in the balancing results 3 weights of the same weight should be present on the grinding wheel which are spaced...

Страница 22: ...ocess can return one step back leave save the result and store parameters Fig 19 a first run b the calculation of weights c the result of balancing 7 4 4 Displaying saved balancing Main menu Measure B...

Страница 23: ...VIBRODIAG www viditech eu sales viditech cz 22 Fig 18 a Saved data b Informative data c Measured results...

Страница 24: ...hreshold absolutely in mm s or relatively by entering percent of range The numerical value of threshold is entered by tapping on the box marked Threshold It is possible to record Velocity analysis Acc...

Страница 25: ...02 30 and for Vibration 00 00 02 fig 21a After analysis initiation the following are displayed time till end of one period and saving of first and subsequently the following record If the symbol is in...

Страница 26: ...nnel sensing are preset Maximal number of analyses that can be loaded for the threshold recorder is 10 only 5 of them are displayed in the graph Information about the current selected analysis are sho...

Страница 27: ...AG www viditech eu sales viditech cz 26 Fig 22 a Saved data from rec b Threshold rec vibrations c Threshold rec analyses Fig 23 a Adding another analysis b Periodical rec vibrations c Periodical rec a...

Отзывы: