background image

1. If the measured voltage is unsure beforehand, should set the range knob to the highest range, 

then, switch to a proper range according to the displayed value. 

2.If LCD displays “1”, it means over range, should set the range knob to a higher range. 

4-4.DCA MEASUREMENT 

4-4-1.Insert the black test lead to “COM” jack and the red one to “mA” jack

max. 200mA

),

or 

insert the red one to “10A” jack

max. 10A

4-4-2.Set  the  range  knob  to  a  proper  DCA  range,  connect  the  test  leads  across  to  the  circuit 

under tested, the current value and polarity of the point which red lead connect will display on 

LCD. 

NOTE:   

1.If the measured current is unsure beforehand, should set the range knob to a higher range, then, 

switch to a proper range according to the displayed value. 

2.If LCD displays “1” , it means overrange, should set the range knob to a higher range. 

3.Max. input current is 200mA or 10A

subject to where red lead insert

),

excessive current will 

blow  the  fuse.  Be  careful  when  measuring  10A  due  to  un-fused.  Continuously  measuring 

large current may heat the circuit, affect the accuracy, even damage the meter. 

4-5. ACA MEASUREMENT 

4-5-1. Insert the black test lead to “COM” jack and the red one to“mA” jack

max. 200mA

),

or 

insert the red one to “10A” jack

max. 10A

4-5-2.Set  the  range  knob  to  a  proper  ACA  range;  connect  the  test  leads  across  to  the  circuit 

under tested. 

NOTE:   

1.If  the  measured  current  range  is  unsure  beforehand,  should  set  the  range  knob  to  the  highest 

range, then set to a proper range according to the displayed value. 

2.If LCD displays “1” , it means overrange, should set the range knob to a higher range. 

3.Max.  input  current  is  200mA  or  10A

subject  to  where  the  red  lead  insert  to

excessive 

current  will  blow  the  fuse.  Be  careful  when  measuring  10A  due  to  un-fused.  Continuously 

measuring large current may heat the circuit, affect the accuracy, even damage the meter. 

4-6. RESISTANCE MEASUREMENT 

4-6-1. Insert the black test lead to “COM” jack and the red one to“V/Ω” jack. 

4-6-2.Set the range knob to a proper resistance range, connect the test leads across to the 

resistance under measured. 

NOTE: 

1.  If  the  resistance  value  being  measured  exceeds  the  max  value  of  the  range  selected,  LCD 

displays "1", thus, should set the range knob to a higher range. When the resistance is over 1MΩ, 

the meter may take a few seconds to stabilize. This is normal for high resistance readings.   

2.When input terminal is in open circuit, overload displays. 

3.When  measuring  in-line  resistance,  be  sure  that  power  is  off  and  all  capacitors  are  released 

completely. 

4-7. CAPACITANCE MEASUREMENT

 

4-7-1.Insert the red test lead to “10A” terminal and the black one to “mA Cx” jack. 

4-7-2.Set  the  range  knob  to  a  proper  capacitance  range,  connect  the  test  leads  to  the  capacitor 

under measured

note: the polarity of red test lead is “+”

NOTE:   

1.f the capacitance range under measured is unsure beforehand, should set the range knob to the 

highest range, then, set to a proper range according to the displayed value. 

2.If LCD displays“1” , it means over range, should set the range knob to a higher range. 

3.Before  measuring,  LCD  display  might  not  be  zero,  the  residual  reading  will  be  decreased 

gradually and could be disregarded. 

4.When measuring large capacitance, if creeps seriously or break capacitance, LCD will display 

some instability value. 

5.Discharge all capacitors completely before capacitance measurement to avoid damage. 

6.UNIT:    1uF =1000nF        1nF=1000pF 

4-9 .TEMPERATURE MEASUREMENT 

            Insert the cathode of thermocouple's cold end to "mA        "jack and anode   

To"COM"terminal, put the working end on or in the tested object, tempe-rature value can be 

read on LCD in Celsius. 

4-9-1.Insert the black test lead to “COM” terminal and the red one to V/Ω jack (Note: the 

polarity of red test lead is “+”). 

4-9-2.Set the range knob to “

” range, connect the test leads to the diode under measured, 

reading is the approximation of the diode positive volt drop. 

4-9-3.Connect  the  test  leads  to  two  points  of  the  measured  circuit,  if  buzzer  sounds,  the 

resistance is lower than approx.(30±20)Ω. 

4-10.DATA HOLD 

Press HOLD key, the present value will be hold on LCD. 

5.MAINTENANCE 

  DO NOT try to verify the circuit for it’s a precision meter. 

5-1.Be ware of waterproof, dustproof and shockproof. 

5-2.Do not operate and store the meter in the circumstance of high temperature, high humidity, 

and flammability, explosive and strong magnetic field. 

5-3.Use the damp cloth and soft solvent to clean the meter, do not use abrasive and alcohol. 

5-4.If do not operate it for a long time, should take out the battery.   

5-4-1.When LCD displays “

” symbol, should replace the battery as below:   

5-4-1-1. Take out the holster and drop out the battery case. 

5-4-1-2  take  out  the  battery  and  replace  a  new  one.  It’s  better  to  use  alkalescency  battery  for 

long time use. 

5-4-1-3.Fix the battery case and take on the holster. 

5-4-2.Fuse replacement 

To use the specified type when replacement. 

5-4-2-1. Take out the holster and drop out the battery case and the battery inside. 

5-4-2-2  When  you  change  the  fuse  of  200mA,  please  replace  a  new  one  as  the  following  fig, 

then fit on the battery case, screw on the case and fit on the holster. 

         

 

 

 

5-4-2-3. When you change the fuse of 10A, please take  off the 9V battery and the screw fit on 

the  back  shell,  then  unscrew  the  fixed  screws  which  showed  on  the  following  fig,  uncover  the 

PCB, and replace a new 10A fuse on the position of the small board showed., finally fit on the 

battery case, screw on the case and fit on the holster. 

(It should operated byprofessional person.) 

 

       

 

6. If the meter does not work properly, check the meter as following:   

CONDITIONS 

WAY TO SOLVE 

NO DISPLAYING 

 

Power is off 

 

HOLD key 

 

Replace battery 

  symbol displays 

 

Replace battery   

NO CURRENT 

INPUT 

 

Replace fuse 

BIG ERROR 

 

  Replace battery 

 

 

 

 

 

 

 

 

 

                                                                                     

 

ET-

 /V1.0 

Отзывы: