 DIGITAL MULTIMETER OPERATION MANUAL 

 

1. INTRODUCTION 

This  instrument  is  a  high  performance  stability,  battery  operated,  3½  digital,  25mm  high 

character jumbo LCD reading to 1999  multi-meter with the function of  measuring DC and 

AC  voltage,  DC  and  AC  current,  Resistance,  temperature  and  Capacitance,  Diode  and 

Continuity  test,  data  hold.  The  Dual-slop  A/D  converter  uses  CMOS  technology  for 

auto-zero, polarity selection and over-range indication. Full overload protection is provided. 

Because of its outstanding features, it is most suitable for use in production line, lab, R & D, 

maintenance and repair work. 

 

2. SAFETY NOTE 

The meter meets the standards of IEC1010. Read the operation manual carefully before operation.

 

2-1

Do not input limit over-ranged. 

2-2

The voltage below 36V is safety. To avoid electric shock, check whether the test leads 

are connected correctly, whether the insulation is good when measuring over 36DCV 

or 25ACV. 

2-3

Remove the test leads when changing function and range. 

2-4

To select correct function and range, beware of error operation.

 

2-5

Do not operate the meter if battery case and back cover is not fixed. 

2-6

Do not input voltage when measuring resistance. 

2-7

Remove test leads from test point and turn off the power before replacing battery and fuse. 

2-8

Safety symbols. 

”Exists dangerous voltage

”    GND

”Dual insulation 

”The operator must refer to the manual

”Low battery 

 

3. CHARACTERISTIC 

3.1 GENERAL 

      3-1-1.    Display: LCD displaying. 

3-1-2.    Max. displaying: 1999

3 1/2digit

auto polarity indication. 

3-1-3.    Measuring method : dual slope A/D conversion. 

3-1-4.    Sampling rate: approx. 3 times/second. 

3-1-5.    Over range indication: the MSD displays“1” or“-1”. 

3-1-6.    Low battery indication:“

” appears. 

3-1-7.    Operation environment: Temperature (0

 to 40

), humidity<85%RH. 

3-1-8.    Storage environment: Temperature (-10

 to 50

), humidity<85%RH. 

3-1-9.    Power

9V×1

NEDA1604/6F22 or equivalent model

3-1-10.    Size: 155mm×90mm×48mm 

3-1-11.    Weight: approx. 270g

including battery

3-1-12.    Accessories:  manual,  holster,  gift  box,  test    lead  a  pair  of  TP01  thermocouple  s

9Vbattery. 

3.2 ELECTRICAL SPECIFICATIONS 

    3-2-1. Accuracy is ±(percentage of r number of digit) at 23±5

,<75%RH. 

    3-3-2.

DCV   

Range 

Accuracy 

Resolution 

200mV 

±(0.5%+3)

 

100uV 

2V 

1mV 

20V 

10mV 

200V 

100mV 

1000V 

±(0.8%+5)

 

1V 

Input impedance: 10MΩ of all ranges. 

Overload protection: 250V DC or AC peak value at 200mV range. 

1000V DC or AC peak value at other ranges. 

 

3-2-3.ACV   

Range 

Accuracy 

Resolution 

20V 

±(0.8%+5)

 

10mV 

200V 

100mV 

750V 

±(1.2%+5)

 

1V   

Input impedance: 10MΩ of all ranges. 

Overload protection: 1000V DC or AC peak value. 

Frequency response:

 ≤

200V range: (40~400)Hz. 

                                            750V range: (40~200)Hz. 

Display: Sine wave RMS (mean value response) 

     

      3-2-4.DCA 

Range 

Accuracy 

Resolution 

20mA 

±(0.8%+5)

 

10uA 

200mA   

100uA   

10A 

±(2.0%+8)

 

10mA 

Max input voltage: 200mV 

Max input current: 10A (up to 10 seconds) 

Overload protection

Fast 0.2A/250V fuse, 10A/250V fuse. 

 

 

  3-2-5.ACA 

Range 

Accuracy 

Resolution 

20mA 

±(1.5%+5)

 

10uA 

200mA   

100uA   

10A 

±(3.0%+10)

 

10mA 

      Max measuring voltage: 200mV 

Max input current: 10A (up to 10 seconds) 

Overload protection

Fast 0.2A/250V fuse, 10A/250V fuse   

Frequency response: (40~200)Hz 

Display: Sine wave RMS (mean value response) 

 

3-2-6.RESISTANCE 

Range 

Accuracy 

Resolution 

200Ω 

±(0.8%+5)

 

0.1Ω 

2kΩ 

±(0.8%+3)

 

1Ω 

20kΩ 

10Ω 

200kΩ 

100Ω 

2MΩ 

1kΩ 

20MΩ 

±(1.2%+8)

 

10kΩ 

Open circuit voltage: less than 3V. 

Overload protection: 250V DC or AC peak value. 

NOTE: at 200Ω range, the test leads should be short-circuit, and measure the down-lead 

resistance, then, subtract from the real measuring. 

WARNING: Do not input any voltage at resistance range for safety! 

          3-2-7.

CAPACITANCE 

Range 

Accuracy 

Resolution 

20nF 

±(3.0%+10)

 

10pF 

2uF   

1nF   

200uF 

100nF 

Overload protection: 36V DC or AC peak value. 

 

 

 

 

3-2-8. TEMPERATURE 

            Range 

Accuracy 

Resolution 

(-40 ~ 1000)

 

< 400

±(0.8%+4) 

≥ 400

±(1.5%+15) 

1

 

Sensor: K-type thermocouple    ( banana shape plug ) 

 

3-2-9.DIODE AND CONTINUITY TEST 

Range 

Description 

Test Condition 

 

  Forward voltage drop 

of diode 

Forward DC Current approx. 

1mA. Reversed DC voltage 

approx. 3V. 

Buzzer sounds if 

resistance 

between terminals 

V/Ωand COM is less 

than about( 30±10)Ω. 

Open circuit voltage approx.    is 

3V 

Overload Protection: 250V DC/AC RMS 

WARNING: Do not input any voltage at resistance range for safety! 

 

     

4. OPERATION 

4-1. FRONT PANEL DESCRIPTION 

1.

 

LCD display: display measuring data and symbol.   

2.

 

DATA  HOLD  switch:    press  this  key,  “H”  will 

display and the present value will be hold on LCD. 

3.

 

FUNCTION  and  RANGE  switch:  select  the 

function and range. 

4.

 

test  jack  of  current  <200mA,capac”  and   
temperature”-“ pole. 

5.

 

test jack of 10A current and capacitance “-“. 

6.

 

voltage and resistance test jack. 

7.

 

GND and temperature “+” test jack. 

8.

 

battery case.   

 

4-2. DC VOLTAGE MEASUREMENT 

4-2- 1. Insert the black test lead to “COM” jack, the red one to V/Ω jack. 

4-2-2. Set the range knob to a proper DCV range, connect the test leads across to the circuit 

under tested, the polarity and voltage of the point which red lead connect will display 

on LCD. 

NOTE:   

1.  If  the  measured  voltage  is  unsure  beforehand,  should  set  the  range  knob  to  the  highest 

range, then, switch to a proper range according to the displayed value. 

2. If LCD displays “1”, it means over range, should set the range knob to a higher range. 

 

4-3.AC VOLTAGE MEASUREMENT 

4-3-1.Insert the black test lead to “COM” jack, the red one to V/Ω jack. 

4-3-2.  Set  the  range  knob  to  a  proper  ACV  range,  connect  the  test  leads  across  to  the  circuit 

under tested. 

NOTE:                       