41
Enter to confirm. Then according to the prompt, enter 20kΩ, then set the true value,
after the measured value is stable press Enter to confirm. Then the calibration for
20kΩrange is completed, and the program will automatically switch to the 200kΩ level,
and change the prompt.
200kΩ: short circuit for the input terminals, after the measured value is stable press
Enter to confirm. Then a
ccording to the prompt, enter 200kΩ, then set the true value,
after the measured value is stable press Enter to confirm. Then the calibration for
200kΩrange is completed, and the program will automatically switch to the 2MΩ level,
and change the prompt.
2M
Ω: short circuit for the input terminals, after the measured value is stable press
Enter to confirm. Then according to the prompt, enter 2MΩ, then set the true value,
after the measured value is stable press Enter to confirm. Then the calibration for
2MΩrange is completed, and the program will automatically switch to the 20MΩ level,
and change the prompt.
20MΩ: short circuit for the input terminals, after the measured value is stable press
Enter to confirm. Then according to the prompt, enter 20MΩ, then set the true value,
after the measured value is stable press Enter to confirm, and the secondary display
shows OK. Then the resistance calibration is complete, and the program will write the
above calibration data into FLASH.
6. Capacitance calibration (the following steps serve only as reference; the value
entered should be in accordance with the prompt on the screen)
Enter the resistance calibration interface, the default range for calibration is 2nF
2nF: open circuit for the input terminals, after the measured value is stable press Enter
to confirm. Then according to the prompt, enter 2nF, then set the true value, after the
measured value is stable press Enter to confirm. Then the calibration for 2nF range is
completed, and the program will automatically switch to the 20nF level, and change
the prompt.
20nF: open circuit for the input terminals, after the measured value is stable press
Enter to confirm. Then according to the prompt, enter 20nF, then set the true value,
after the measured value is stable press Enter to confirm. Then the calibration for