IND110 Technical Manual 

5-9

 

Calibration 

Calibration is done with DIP switches 1-4 and the three buttons. For normal 
weighing all the DIP switches are OFF. For calibration, DIP switches 1-4 are 
switched ON as required to calibrate the weigh mode ZERO, weigh mode SPAN, 
LO setpoint cutoff and HI setpoint cutoff. 
 

Set Weigh Mode Zero 
Unload scale. Set DIP switch 1 to ON (all other switches OFF). HI LED will flash 
(coarse setting). With a calibration current meter connected to the output, use “+” 
and “-“ buttons to adjust output current to desired “zero load” value (Typically 4 
mA). If fine adjustment is required, depress  “

4

 

“ button until LO LED flashes. Use 

the “+” and “-“ buttons for fine adjustment. When correct zero value is obtained, 
return DIP switch 1 to OFF position. The zero load current output is stored. 

 

Set Weigh Mode Span 
Load scale with the maximum load weight corresponding to the desired current 
output (Typically 20 mA). Set DIP switch 2 to ON (all other switches OFF). HI LED 
will flash (coarse setting). use “+” and “-“ buttons to adjust output current to 
desired “full load” value (Typically 20 mA). If fine adjustment is required, depress 

4

“ button until LO LED flashes. Use the “+” and “-“ buttons for fine adjustment. 

When correct span value is obtained, return DIP switch 2 to OFF position. The full 
load current output is stored. NOTE: It is possible to calibrate with a value less 
than full load weight. Simply calculate current output value needed to represent the 
load applied and adjust current output for that value. However, the lower that value 
is, the lower the setting accuracy obtained. Always try to use the highest 
calibration weight possible. 

Set HI setpoint cutoff value. 
Load scale with a value equal to the desired HI setpoint cutoff point and note the 
4/20 mA current output reading. Alternatively, calculate the output current at the 
desired load cutoff point. Remember the value and turn ON DIP switch 3 (all other 
switches OFF). HI LED will flash. Using the “+” and “-“ buttons, depress them until 
the 4/20 mA output matches the remembered value. As with the weigh mode 
calibration, the “

 

“ button can be used to toggle into fine adjustment mode if 

needed. Return DIP switch 3 to OFF when complete. 

 

Set LO setpoint cutoff value. 
This setpoint is identical to the HI setpoint calibration except DIP switch 4 is used 
instead. Note: there is no significance to the HI and LO setpoint names. The HI can 
be set for the lowest value if desired. It is simply a way to keep track of the values. 
Also, it is not necessary to use either setpoint. Do not program if not needed.

Содержание IND110

Страница 1: ...IND110 Load Cell Signal Converter Technical Manual 17041600A 5 04 00...

Страница 2: ......

Страница 3: ...ORE CONNECTING DISCONNECTING ANY INTERNAL ELECTRONIC COMPONENTS OR INTERCONNECTING WIRING BETWEEN ELECTRONIC EQUIPMENT ALWAYS REMOVE POWER AND WAIT AT LEAST THIRTY 30 SECONDS BEFORE ANY CONNECTIONS OR...

Страница 4: ...ates is in conformity with the following standard s or other normative document s auf das sich diese Erkl rung bezieht mitder den folgenden Norm en oder Richtlinie n bereinstimmt Auquel se r f re cett...

Страница 5: ...shed with Restricted Rights Copyright 2004 METTLER TOLEDO This documentation contains proprietary information of METTLER TOLEDO It may not be copied in whole or in part without the express written con...

Страница 6: ...terference received including interference that may cause undesired operation This equipment has been tested and found to comply with the limits for a Class A digital device pursuant to Part 15 of FCC...

Страница 7: ...For your notes...

Страница 8: ...Chapter 3 0 Specifications 3 1 Chapter 4 0 Dimensions 4 1 Chapter 5 0 Installation 5 1 Load cell connections 5 1 Load cell Selection 5 2 Module mounting details 5 3 Analog Output 4 20 mA Connections 5...

Страница 9: ...it and then converting it back to an analog 4 20 mA signal It would then be typically sent to a remote analog input such as a PLC In addition to the 4 20 mA output signal two discrete setpoint output...

Страница 10: ...IND110 Technical Manual 1 2 For your notes...

Страница 11: ...del of the IND110 Signal Converter The IND110 is not suitable for mounting into a harsh environment without a protective enclosure An optional stainless IP66 rated enclosure with gasket is available s...

Страница 12: ...of enclosure is available when installation in hazardous areas is required by ordering part number 17041800A It is rated by the manufacturer for use in a Class I Group C D Class II Group E F G Class I...

Страница 13: ...put Resolution 1 part in 4000 min 25 cell span Output Linearity 0 012 max Temp Stability 20 PPM C span 50 PPM C zero Conversion time 50 mS 20 updates sec Isolation Circuit common connected to supply c...

Страница 14: ...IND110 Technical Manual 3 2 For your notes...

Страница 15: ...Chapter 4 0 Dimensions METTLER TOLEDO IND110 PWR Lo Hi 5 57 7 7 74 4 1 10 00 0 Module Dimensions Figure 4 4 0 101 6 7 5 190 5 2 75 69 9 6 25 158 75 7 0 177 8 Stainless Steel Enclosure Dimensions Figur...

Страница 16: ...IND110 Technical Manual 4 2 Explosion Proof Enclosure Dimensions Figure 6 7 75 196 9 7 0 177 8 6 0 152 4 8 0 203 2 5 0 127 6 75 171 5...

Страница 17: ...C SAFE BARRIER FAILURE TO OBSERVE THESE PRECAUTIONS COULD RESULT IN BODILY HARM AND OR PROPERTY DAMAGE Load cell connections must be made with shielded cable as shown below Note that the IND110 SHIELD...

Страница 18: ...west resistance allowed 3 Calculate the full scale load cell signal output a Mv system capacity x single cell output rating x excitation voltage single load cell capacity x number of cells weighing ra...

Страница 19: ...has two tapped 3 4 NPT holes only Sealed explosion proof fittings are required to complete the installation The installer must provide them insure they are properly installed and sealed according to...

Страница 20: ...ns snap the module onto the DIN rail first and then install into the enclosure First install the M6 screws loosely into predrilled holes in the enclosure Next place module and rail onto screws The rai...

Страница 21: ...on and 4 20 mA common lines together As a result connect only one of these points to an earth ground Many process control instruments connect their 4 20 mA common to earth internally If true this syst...

Страница 22: ...ngerous The supply chosen must also have an isolated output This is important because the 24 VDC common load cell shield setpoint common and the 4 20 mA common are internally connected together These...

Страница 23: ...nsistor can sink 75 mA drive a load rated at 75 mA or less The load must be connected to an ungrounded voltage of 24VDC or less CAUTION If the user s DC load supply is grounded some form of isolation...

Страница 24: ...n Off Off Off Off Off flash off Fine zero adjustment mode off flash Coarse span adjustment mode Span Calibration Off On Off Off Off Off flash off Fine span adjustment mode off flash Coarse HI setpoint...

Страница 25: ...for fine adjustment When correct span value is obtained return DIP switch 2 to OFF position The full load current output is stored NOTE It is possible to calibrate with a value less than full load we...

Страница 26: ......

Страница 27: ...D110 make sure the power has been removed There are no field serviceable parts within the IND110 WARNING DISCONNECT ALL POWER TO THIS UNIT BEFORE INSTALLING SERVICING OR CLEANING FAILURE TO DO SO COUL...

Страница 28: ...6 2 For your notes...

Страница 29: ......

Страница 30: ...1900 Polaris Parkway Columbus Ohio 43240 www mt com P N 17041600A METTLER TOLEDO is a registered trademark of Mettler Toledo Inc 2004 Mettler Toledo Inc Printed in USA 17141600A...

Отзывы: