NM PLUS HYDROGEN GENERATOR
USER MANUAL
www.vicidbs.com
Version 1.06.0000
Lock Keyboard
Keyboard will be locked after the generator
is in the main window for more than 20s. To
unlock, press the unlock key and hold for 5s
Yes/No
No
Set Default Values
Sets all configuration parameters to default
7.2.2. Diagnostic Display
ITEM
DESCRIPTION
Production Tot.
Total production of hydrogen
Operating Time (h)
Total number of hours the unit is operational
Wat. Quality (μS)
Actual water conductivity
Int. Press.
Actual internal pressure of the unit
Cell Voltage (V)
Actual cell voltage
Cell Current (A)
Actual cell current
PS. Temp.
Actual temperature of the main board
Cell Voltage Peak (V)
The maximum voltage the cell ever reached
PS. Temp. Peak
The maximum temperature the main board reached
Pump Flow (L/min)
Actual flow of water
7.2.3. Special Functions
ITEM
DESCRIPTION
Dryer Status
Shows the temperature of the drying columns and the position of the cycle
System Test
This offers several test functions - see below
Internal Leak Test
The outlet valve is closed, the pressure is set to the max., when the pressure
reaches the max., production is stopped and the pressure drop is measured
during 2 minutes. The pressure drop should be below 0.7 bars or 10 psi.
Important: The leak test will only work if the generator is in standby mode.
External Leak Test
Works similar to the internal leak test except that the outlet valve is open
and the test pressure is equal to the set pressure.
This function can be used to test the external gas lines
Important: The external leak test will only work if the generator is in standby
mode.
Flow Test
This function sets the outlet valve to provide a certain flow
adjustable from 0 to 100% of maximum flow.
Important: This function will only work if the generator is in standby mode. It
will take approx. 2 minutes to get a stable flow!
Rinse Flow Test
This test checks the gas flow for the dryer regeneration
Complete System Test
Combination of all tests above
Hardware Options
Not available for user (password protected) Contact dedicated service
organization