65
8-4-1(2) Span test with external weight
Make sure to use the external weight which is equal to the weighing capacity of each model.
1
Select the span test with external weight.
- Press [Menu] key, then press [1]-[4] keys
to go to <632 EX SPAN TEST>.
- Press [4] key to execute.
2
Zero-point adjustment starts.
Display changes to the order of "blinking of
<buSY>"→ “blinking of <on 0>”.
On completion of the zero-point
adjustment, the display automatically
changes to <on FS>.
3
The span test starts.
For HJ17K0.1S(R) and HJ22K0.1S(R):
For HJ33K0.1S(R), and HJ62K0.1DS(R):
- Place the weight in the centre of the
weighing pan.
・
For HJ17K0.1S(R) and HJ22K0.1S(R):
Display changes to the order of
<on FS> → “blinking of <on FS>”.
Then span test starts.
・
For HJ33K0.1S(R), and HJ62K0.1DS(R):
Display changes to the order of
<on FS> → <push M>.
- Press [Menu] key.
Display changes to “blinking of
<on FS>”, then span test starts.
On completion of the span test, the display
automatically changes to
<INSTRUMENTAL ER> and the
instrumental error of the scale is displayed.
- Press [4] key.
<632 EX SPAN TEST> is displayed.
- Unload the weight from the weighing pan.
4
Exit the setting menu.
- Press [Menu] key to shift to the measuring
mode.