4 O
PERATION
Pass/Fail mode
An integrated pass/fail analysis feature simplifies standard
conformity testing and provides unambiguous measurement
result presentation.
The OLP
-
37XV2 provides two Pass/Fail modes:
• PF Abs: Pass/Fail indication based on a absolute power
threshold
• PF Loss: Pass/Fail indication based on a loss threshold
To select a Pass/Fail mode:
J
Press [
] until PF Abs or PF Loss is displayed.
NOTE:
When PF Abs or PF Loss is selected the two modes can also be
toggled by pressing [
].
Setting the fail threshold
Setting the fail threshold is identical for both Pass/Fail modes.
I
PF Abs or PF Loss mode is selected. Press [
] to toggle
between both modes.
1.
Press [
] to select a wavelength (see also page
).
2.
Long press [
].
The Set Fail Threshold screen opens.
3.
Use [+]/[–] to change the threshold.
4.
Press [
] to save the new threshold.
NOTE:
When Dual mode is active, the fail threshold can not be set.
Select G-PON or XGS-PON to set threshhold individually.