OLP
-
35V2/
-
35SC/
-
38V2
23
P
ASS
/F
AIL
MODE
4 O
PERATION
Pass/Fail mode
An integrated pass/fail analysis feature simplifies standard
conformity testing and provides unambiguous measurement
result presentation.
The OLP-3xV2 provides two Pass/Fail modes:
• PF Abs: Pass/Fail indication based on a absolute power
threshold
• PF Loss: Pass/Fail indication based on a loss threshold
To select a Pass/Fail mode:
J
Press [
] until PF Abs or PF Loss is displayed.
NOTE:
When PF Abs or PF Loss is selected the two modes can also be
toggled by pressing [
].
Setting the fail threshold
Setting the fail threshold is identical for both Pass/Fail modes.
I
PF Abs or PF Loss mode is selected. Press [
] to toggle
between both modes.
1.
Press [
] to select a wavelength (see also page
19).
2.
Long press [
].
The Set Fail Threshold screen opens.
3.
Use [+]/[–] to change the threshold.
4.
Press [
] to save the new threshold.
NOTE:
When Multi-
is active, the fail threshold can not be set.
To deactivate Multi-
e.g. switch of the laser or disable Multi-
at
the transmitter side temporarily.