Chapter 2
Setting up the system
Updating the system software
ONA-800 Getting Started Guide
Page 24
22142812, R011
April 2022
5
Using Windows Explorer, navigate to the software file, double-click the file, and
then click
Run
.
6
In the dialog that appears, click
Browse
to navigate to and select the USB drive.
7
Click
Extract
to initiate extraction of the system software files to the USB drive.
Do not unplug the drive from the PC while the files are extracting.
8
When all the files are extracted, navigate to the USB drive, right-click the name of
the drive, and then click
Eject
.
9
Unplug the USB drive from the PC.
10
If required, go to
“Performing a software update using a USB drive”
.
Performing a software update using a USB drive
1
Confirm the following:
–
The unit will have an uninterrupted supply of power during the software
update. If necessary, connect the AC power adapter to the unit.
–
A USB drive on which the extracted system software files are saved is
available.
2
Plug the USB drive into the USB connector on the top panel of the unit.
3
Select
System
on the menu bar, select
Upgrade
, and then select
USB
.
The release number of the software available on the USB drive appears. Note the
following symbols that may appear beside the release number on the USB stick:
–
An equals symbol indicates that the software on the USB stick and the
software on installed on the unit are the same (that is, have the same release
number).
–
A down-arrow symbol indicates that the software on the USB stick is an
earlier release than the software installed on the unit.
–
An up-arrow symbol indicates that the software on the USB stick is a later
release than the software installed on the unit and the system software can
be upgraded to the later release.
4
To upgrade to a later release, select
Start Upgrade
.
A dialog box appears with prompts to either exit all tests that are running or cancel
the update if the tests cannot be stopped.
5
Select
OK
to proceed with the software update. Do not unplug the USB drive from
the unit during the update.
The unit automatically restarts when the update is completed.
6
After the unit has restarted, unplug the USB drive.
WARNING:
Electrical shock may result in serious injury or death. Be sure the AC power adapter
is connected to the correct voltage mains. Do not use the adapter outdoors or in wet
locations. Use only the AC power adapter supplied with the unit.