background image

 

 

 

 

VIAVI Solutions

 

Quick Card 

 

OneAdvisor-800 Modular Test Set

  

Measuring Optical Power Levels with a 4100-series OTDR Module 

 

 
This procedure describes how to use an E4100-series OTDR module as an Optical Power Meter (OPM).  
If optioned, the OTDR module can operate as an OPM or Optical Light Source, but not at the same 
time.  Light Source is a standard feature on most new OTDR modules, while Power Meter is an option.   
 

Equipment Requirements: 

 

OneAdvisor-800 equipped with the 
following: 

o

 

Fiber Optics Software Release V19.92 
or greater 

o

 

E4100 Series OTDR Module 

o

 

E41OTDRPM Power Meter option  

 

Fiber optic cleaning and inspection tools 

 

Reference Cable (patch cord) with 
connectors matching the OTDR port and 
Fiber Under Test 

 

Optical Coupler to connect Reference 
Cable to Fiber Under Test 

 

 

Figure 1: Equipment Requirements 

The following information is required to complete the test: 

 

Type of Fiber (Multimode or Single Mode) 

 

Type of Connectors (SC UPC, SC APC, LC UPC, etc.) 

 

Wavelength(s) to be tested (850nm, 1300nm, 1310nm, 1550nm, etc.) 

 

Fiber Inspection Guidelines: 

 

Use the VIAVI P5000i or FiberChek Probe 
microscope to inspect both sides of 
every connection being used (OTDR Port, 
Reference Cable, bulkhead connectors, 
patch cords, etc.) 

 

Focus the fiber on the screen.  If dirty, 
clean the end-face. 

 

If it appears clean, run inspection test. 

 

If it fails, clean the fiber and re-run 
inspection test.  Repeat until it passes.

 

 

 

Figure 2: Inspect Before You Connect (IBYC) 

Отзывы: