FVAm™ Series Benchtop MPO Autofocus Microscope User Guide
22132675 Rev. 101, Standard, 3-8
Performing Pass/Fail tests and working with saved test results
Editing the active optical setting
Step 4 To adjust the sizing, position the blue circle until it is just inside the fiber edge.
Press the
Manual Focus/Navigation
button up or down, or move the touchscreen slider to the left or
right.
Step 5 Select
NEXT
.
Step 6 To adjust the lighting, do one of the following:
▪ Select
AUTO
to perform an automated brightness adjustment.
▪ Manually adjust the lighting, and then select
NEXT
.
Press the
Manual Focus/Navigation
button up or down, or move the touchscreen slider to the left or
right.
Step 7 Select
DONE
, and then select
EXIT
.
— End —