8 Electromagnetic compatibility
The device meets the requirements of IEC 60601-1-2.
Warnings and Cautions
•
This device should not be used in the vicinity or on the top
of other electronic equipment such as cell phone, transceiver
or radio control products. If you have to do so, the device
should be observed to verify normal operation.
• The use of accessories and power cord other than those
specified, with the exception of cables sold by the
manufacturer of the equipment or system as replacement
parts for internal components, may result in increased
emissions or decreased immunity of the equipment or
system.
Guidance and manufacturer
’
s declaration
–
electromagnetic emissions
The model Dynamic ECG recorder is intended for use in the electromagnetic
environment specified below. The customer or the user of the model Dynamic
ECG recorder should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions CISPR 11
Group 1
The model Dynamic ECG recorder uses RF
energy only for its internal function. Therefore,
its RF emissions are very low and are not likely
to cause any interference in nearby electronic
equipment.
RF emissions CISPR 11
Class B
The model Dynamic ECG recorder is suitable
for use in all establishments, including
domestic establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings used
for domestic purposes.
Harmonic emissions
IEC 61000-3-2
n.a.
Voltage fluctuations/
flicker emissions IEC
61000-3-3