14
Product Compatibility
The
Elite
was not designed to be compatible with other products. VIASYS Healthcare
recommends using only VIASYS approved accessories for use with the Elite.
Warning: No part of the
Elite
is to be connected to a telephone system.
Warning: This equipment is not intended for use with HF surgical equipment.
SPECIFICATIONS
Dimensions:
17cm x 7.5cm x 2.5cm (6.8” x 2.9” x 1.0”)
Weight (with one probe):
Approximately 0.25 kg (.60 lb.)
Doppler Technology:
Continuous Wave (CW) Unfocused.
Battery Type:
9 volt alkaline or 9 volt rechargeable [IEC #6LR61 or
NEMA # 1604A]
Operational Temperature:
10 to 32 degrees C (50 to 90 degrees F)
Speaker:
2.25", 8 Ohms
Audio Output
:
1/4 watt typical
Headphones:
8 Ohms, mono 3.5mm
Frequency Response:
-3dB @ 360 Hz and 1.2 K for
Elite
100 Series
-3dB @ 391 Hz and 1.2 K for
Elite
200 Series
Safety Standards:
IEC 601-1 and IEC601-1-2 (if applicable and only when using
the appropriate recharger)
IEC 60601-2-37
Classification:
Internally Powered
Equipment, Type B
Display Rate (BPM):
60-119 ± 2 for Vascular Probe
(If applicable)
60-199 ± 2 for OB Probe
Recharger:
Input:
120 VAC 60 Hz (PN: PS-0019 )
or
240VAC 50Hz @ 6VA (PN: PS-0028)
Output:
24VDC @ 65ma
The maximum surface temperatures of the ultrasonic transducers when applied to the patient are as follows:
2 MHz:
24.6 degrees C
2 MHz waterproof:
23.5 deg. C
3 MHz:
24.4 deg. C
5 MHz:
27.6 deg. C
8 MHz:
27.7 deg. C