Страница 2: ...and Indicators 5 2 7 Alternating Between Display Modes short button press 6 2 8 Adjusting Display Brightness long button press 6 3 Error Messages Problems Corrective Actions 7 3 1 General Information 7 3 2 Error Messages Causes 7 3 3 Problems Causes Corrective Actions 7 3 4 EMI Electromagnetic Interference 7 4 Maintenance Cleaning 8 5 Symbol Definitions 8 6 Technical Specifications 8 7 Packing List ...

Страница 3: ...nts and pulse signals can be affected by certain environmental conditions oximeter application errors and certain patient condi tions See the appropriate sections of this manual for specific safety information If you are uncertain about the accuracy of any measurement check the patient s vital signs by alter nate means then ensure that the VM 2101 is functio ning correctly The VM 2101 is not defibri...

Страница 4: ...r nail facing the top of the finger oximeter so that the finger is placed fully on the silicone pad Release the nipped end of the finger oximeter to secure it on to the patient s finger 2 3 Inserting the Finger 2 2 Switching on the Device Press and hold the button on the front panel briefly until the opening welcome screen appears After the power on self test is successfully completed the device is rea...

Страница 5: ... weak pulses The colour of the bar graph is an indicator for signal quality Green good signal quality very accurate measurement Yellow average signal quality measurement may be inaccurate Red poor signal quality unreliable measurement 4 Pulse rate in beats per minute 5 Pleth waveform The reading is automatically adjusted to the pulse strength therefore a waveform with strong amplitude should be vi...

Страница 6: ...edium or low brightness levels and a more frequent change of batteries may be required The finger oximeter has 6 levels of adjustable bright ness If whilst in operation the button is kept pressed for longer than one second the brightness adjustment be comes active and the brightness adjustment screen is shown For every second that the button is kept pressed the brightness increases by one level Onc...

Страница 7: ... activity may be preventing the monitor from tracking the pulse Keep the patient still if possible Change the appli cation site There may be interference due to ambient light or the finger oximeter may be on an extremity with a blood pressure cuff arterial catheter or intravascular line Reposition the finger oximeter as necessary Electromagnetic interference may be preventing the monitor from tracki...

Страница 8: ...nger oximeter Attention See instructions for use Manufacturer Date of manufacture Type BF S N Serial number P N Part number Observe applicable waste disposal regulations European Union approval 5 Symbol Definitions 6 Technical Specifications Measurement Range SpO2 0 to 100 Pulse Rate 20 to 300 beats per minute bpm Accuracy SpO2 2 70 to 100 Pulse Rate 1 digit 100 bpm 1 100 bpm Display OLED colour gra...