SPeCIfICATIoNS ANd dISPlAYed eRRoR CodeS
GeNeRAl SPeCIfICATIoNS
mODEL
CAPACITY /
grADUATION
WEIgHT
(Kg)
DImENSION mm
BATTErY
VB2-30-EC
max 30 kg / 50 g
3,6
210 x 260x 65
6 of alk. 1,5 V batteries lR6/AA for 15 hrs operation
VB2-100-EC
max 100 kg /10 (50) g
4,5
260 x 345 x 55
6 st alk. 1,5 V batteries lR6/AA for 24 hrs. operation
WS-100-EC
max 100 kg / 50 g
4,5
260 x 345 x 55
6 of alk. 1,5 V batteries lR6/AA for 15 hrs operation
VB2-200-EC
max 200 kg / 100 g
4,5
260 x 345 x 55
6 of alk. 1,5 V batteries lR6/AA for 15 hrs. operation
VB2-30-05
max 30 kg / 5 g
Not approved version
3,6
210 x 260x 65
6 of alk. 1,5 V batteries lR6/AA for 15 hrs. operation
VB2-100-02
max 100 kg / 2 g
4,5
260 x 345 x 55
6 of alk. 1,5 V batteries lR6/AA for 15 hrs. operation
VB2-100-10
max 100 kg / 10 g
Not approved version
4,5
260 x 345 x 55
6 of alk. 1,5 V batteries lR6/AA for 15 hrs. operation
VB2-130-10
WS-130
max 130 kg / 10 g
Not approved version
4,5
260 x 345 x 55
6 of alk. 1,5 V batteries lR6/AA for 15 hrs. operation
VB2-100-50
max 100 kg / 50 g
Not approved version
4,5
260 x 345 x 55
6 of alk. 1,5 V batteries lR6/AA for 15 hrs. operation
VB2-200-10
max 200 kg / 20 g
Not approved version
4,5
260 x 345 x 55
6 of alk. 1,5 V batteries lR6/AA for 15 hrs. operation
- Alternative power is a 9 VdC adaptor (centre minus).
OPErATOr INTErFACE
Display
0.75” (19 mm) 7-segment, liquid Crystal, 6-digit
Additional Symbols
Net, Gross, Stable, Tare, lb, kg, Zero
Keyboard
5-key flat membrane panel
ENVIrONmENTAL
Operating Temperature
0 to +30 grad C (-10 to +40 grad C not approved version)
Storage Temperature
-25 to +70 grad C
dISPlAYed eRRoR CodeS
Code
mode
meANING / PoSSIBle SolUTIoN
¤¤¤¤
Normal operating m. Gross overload. A weight greater than the rated capacity has been applied to the scale. Remove the weight from
the platter or try re-calibrating the scale. otherwise, check for a bad load cell connection or possible load cell
damage due to overloading.
bAtt
Normal operating m. Indicates a low battery condition. Re-charge the battery or replace if alkaline batteries.
Err 0
Span Calibration
mode (f17)
keyed-in weight value is larger than full-scale capacity. Use a smaller test weight or check keyed-in value.
Err 1
Span Calibration
mode (f17)
keyed-in weight value is less than 1% of full-scale capacity. Use a larger test weight or check keyed-in value.
Err 2
Span Calibration
mode (f17)
There is not enough load cell signal to produce the internal counts necessary to properly calibrate the scale. first check
all load connections. Use f16 mode to view internal counts. See Appendix C for more info.
Err 3
All modes
Non-volatile memory read error. one or more setup parameters have been lost.
Err 4
All modes
Non-volatile memory writes error. Indicator needs service.
Err 5
key-in Span Calibration
mode (f20)
You have attempted to enter a zero value for C1. enter a known calibration value greater than zero.
Err 7
Initialization
No reading from the AdC. make sure there is a load cell(s) connected to the indicator at start-up.
Err 9
Normal operating
Span calibration value has been lost. Re-calibrate the Scale.