SPeCIfICATIoNS ANd dISPlAYed eRRoR CodeS

GeNeRAl SPeCIfICATIoNS

mODEL

CAPACITY  / 

grADUATION

 WEIgHT

(Kg)

DImENSION mm

BATTErY

VB2-30-EC

max 30 kg / 50 g

3,6

210 x 260x 65

6 of alk. 1,5 V batteries lR6/AA for 15 hrs operation

VB2-100-EC

max 100 kg /10 (50) g

4,5

260 x 345 x 55

6 st alk. 1,5 V batteries lR6/AA for 24 hrs. operation

WS-100-EC

max 100 kg / 50 g

4,5

260 x 345 x 55

6 of alk. 1,5 V batteries lR6/AA for 15 hrs operation

VB2-200-EC

max 200 kg / 100 g

4,5

260 x 345 x 55

6 of alk. 1,5 V batteries lR6/AA for 15 hrs. operation

VB2-30-05

max 30 kg / 5 g

Not approved version

3,6

210 x 260x 65

6 of alk. 1,5 V batteries lR6/AA for 15 hrs. operation

VB2-100-02

max 100 kg / 2 g

4,5

260 x 345 x 55

6 of alk. 1,5 V batteries lR6/AA for 15 hrs. operation

VB2-100-10

max 100 kg / 10 g

Not approved version

4,5

260 x 345 x 55

6 of alk. 1,5 V batteries lR6/AA for 15 hrs. operation

VB2-130-10

WS-130

max 130 kg / 10 g

Not approved version

4,5

260 x 345 x 55

6 of alk. 1,5 V batteries lR6/AA for 15 hrs. operation

VB2-100-50

max 100 kg / 50 g

Not approved version

4,5

260 x 345 x 55

6 of alk. 1,5 V batteries lR6/AA for 15 hrs. operation

VB2-200-10

max 200 kg / 20 g

Not approved version

4,5

260 x 345 x 55

6 of alk. 1,5 V batteries lR6/AA for 15 hrs. operation

- Alternative power is a 9 VdC adaptor (centre minus).

OPErATOr INTErFACE

Display

0.75” (19 mm) 7-segment, liquid Crystal, 6-digit

Additional Symbols

Net, Gross, Stable, Tare, lb, kg, Zero

Keyboard

5-key flat membrane panel

ENVIrONmENTAL

Operating Temperature

0 to +30 grad C (-10 to +40 grad C not approved version)

Storage Temperature

-25 to +70 grad C

dISPlAYed eRRoR CodeS

Code 

mode

meANING / PoSSIBle SolUTIoN

¤¤¤¤

Normal operating m. Gross overload. A weight greater than the rated capacity has been applied to the scale. Remove the weight from 

the platter or try re-calibrating the scale. otherwise, check for a bad load cell connection or possible load cell 

damage due to overloading.

bAtt

Normal operating m. Indicates a low battery condition. Re-charge the battery or replace if alkaline batteries.

Err 0

Span Calibration 

mode (f17)

keyed-in weight value is larger than full-scale capacity. Use a smaller test weight or check keyed-in value.

Err 1

Span Calibration 

mode (f17)

keyed-in weight value is less than 1% of full-scale capacity. Use a larger test weight or check keyed-in value.

Err 2 

Span Calibration 

mode (f17)

There is not enough load cell signal to produce the internal counts necessary to properly calibrate the scale. first check 

all load connections. Use f16 mode to view internal counts. See Appendix C for more info. 

Err 3

All modes

Non-volatile memory read error. one or more setup parameters have been lost.

Err 4

All modes

Non-volatile memory writes error. Indicator needs service.

Err 5

key-in Span Calibration 

mode (f20)

You have attempted to enter a zero value for C1. enter a known calibration value greater than zero.

Err 7

Initialization

No reading from the AdC. make sure there is a load cell(s) connected to the indicator at start-up.

Err 9

Normal operating 

Span calibration value has been lost. Re-calibrate the Scale.

Содержание Uniscale 100kg/2g

Страница 1: ...NISCALE200 kg 50 g 200kg 50g Max Overview different ranges 1 Scandinavia s largest webshop for scales Hantverksvägen 15 76493 Vaddo Sweden Tel 46 0 176 208920 Web www vetek com Email info vetek com Weighingblock VB2 series and Uniscale The Weighingblock is a robust scale It is designed for portable use When transported by car between different working places we recommend using of a protected space...

Страница 2: ...hen using a Non Automatic scale in commercial purpose in any of the EC countries it is prescribed by law that the scale must be EC type approved If the purpose of use is not commercial or if the scale s intended use is as an automatic batching scale then there are no legal demands If the Scale is used as a Batching Scale it is therefore an Automatic weighing instruments and there are no legal dema...

Страница 3: ...n for 3 seconds the Scale switches off Used to establish a Tare zero the Scale while in either Gross mode or Net mode This operation cannot be performed at or below Gross zero Configuration Setup menu To do this configuration you require a good acknowledge of how a digital electronic scale works Note If the scale fails it is not always certain the scale needs configuration Excluding the Normal Men...

Страница 4: ...arge the battery or replace if alkaline batteries Indicates stable weighing SETUP MENU 1 Turn the power OFF 2 Put Calibration Switch to the Calibration position 3 Turn the indicator ON 4 The display shows F1 to indicate that the unit is in Setup menu mode 5 To move to a new F heading use TARE or ON to move up or down 6 To move to the selection value press ZERO once 7 Increase and decrease the valu...

Страница 5: ...h determines the point at which the indicator shows overload All selections are based on the primary unit selected in F8 FS Full scale in primary units FS FS 2 all none EC ver FS 1d FS 9d all EC ver F8 Calib Unit Selects the primary base unit to be used in the calibration process Also the default unit for normal operation 1 primary unit is lb 2 primary unit is in kg 1 2 all versions F9 Display Div...

Страница 6: ...17 span adjusts procedure 1 Scroll to F 17 then press ZERO to enter span calibration menu 2 The display will momentarily show C 1 for the span calibration followed by a value with one flashing digit This value will be zero Place the test weight on the Scale 3 Pressing TARE or ON will change the position of the flashing digit 4 Increase the flashing digit by pressing Dot Decrease the flashing digit...

Страница 7: ...Stable Tare lb kg Zero Keyboard 5 key flat membrane panel ENVIRONMENTAL Operating Temperature 0 to 30 grad C 10 to 40 grad C not approved version Storage Temperature 25 to 70 grad C Displayed error codes CODE MODE MEANING POSSIBLE SOLUTION Normal Operating M Gross Overload A weight greater than the rated capacity has been applied to the scale Remove the weight from the platter or try re calibratin...

Страница 8: ...mode 2 Reset to normal mode Check the Loadcell Check the Loadcell with a universal volt instrument see table The left pin is number 1 Terminal J1 Description Colours Loadcell AG Colours Loadcell 1042 Colours Loadcell LOC o FAS Colours Lastcell SP4 E 1 Out 5 VDC Brown Green Red White S 2 In Yellow Red Green Green E 3 Out 0 VDC Green Black Black Black S 4 In White White White Red if it is problem to...

Страница 9: ...mmend visiting the concerning Authorities web sites Enunciator Definitions Pictured above is the LCD display Before reading the value the stable indicator must be on If not wait a few seconds until it appears Handling 1 Press ON to start the scale 2 If necessary press ZERO to obtain a weight reading of zero 3 Press OFF to switch off the scale Hold 3 seconds 4 The scale switches off automatically a...

Страница 10: ...e Scale to be on the floor you can keep it in your hands Start the Scale with ON and wait a few seconds 3 To set the Batching Weight 4 Press Dot display shows 0 0 flashing 5 Press ON until the first relevant digit flash 6 Change to the wished value with Dot 7 Press ON until the next relevant digit flash 8 Change to the wished value with Dot 9 When the Scale shows wished value save with NET 10 Pres...

Отзывы: