Vernier RH-BTA Скачать руководство пользователя страница 2

 

3

How the Relative Humidity Sensor Works 

The heart of this sensor is the Hy-Cal Engineering IH-3602-L Integrated Circuit 

Humidity Sensor, which uses a capacitive polymer to sense humidity. An integrated 

circuit then produces an output voltage which varies with relative humidity. The 

response time of the unit in moving air is much shorter than in still air. In some 

cases, you may want to create air currents (by moving the sensor or using a fan) to 

speed up the response of the sensor. 

The sensor is slightly light sensitive. The housing is designed to minimize the 

amount of light that can penetrate the sensor opening. The calibration of the sensor is 

somewhat affected by temperature. This effect is negligible at the low relative 

humidity readings, but increases at high humidities. If you want to correct for this 

error, you can create different calibration files for different temperatures. In most 

cases, this is unnecessary.  
This sensor is equipped with circuitry that supports auto-ID. When used with 

LabQuest 2, LabQuest, LabQuest Mini, LabPro, Go! Link, SensorDAQ, EasyLink, 

TI-Nspire™ Lab Cradle, or CBL 2™, the data-collection software identifies the 

sensor and uses pre-defined parameters to configure an experiment appropriate to the 
recognized sensor. 

Optional Calibration Procedure 

You do not need to perform a new calibration on the Relative Humidity Sensor. We 

have set each Relative Humidity Sensor to match the stored calibration before 

shipping it. You can simply use the appropriate calibration file that is stored in your 

data-collection program from Vernier.  
For greatest accuracy, this Relative Humidity Sensor can be calibrated. Calibration 

can be done by comparison to another instrument that measures relative humidity 

(hygrometer or psychrometer). Another way is to use salt solutions. You would do a 

two-point calibration as you probably have done with other sensors such as pH 

sensors. The salt solutions are used to maintain environments with different known 

relative humidities. If you place moist salts in a sealed container, the air above it will 

reach a known relative humidity. A table of values is shown on the next page. The 

relative humidity above the salt depends slightly on the temperature, so the table also 

lists temperature.  
Here is a step-by-step procedure, 
1.  Place a handful of salt in the bottom of a jar (quart or liter size is fine). 
2.  Add a little water to the jar so that the salt is wet. The goal is to end up with wet 

salt, not to totally dissolve the salt. 

3.  Place the Relative Humidity Sensor in a jar. Do not get the salt or salt solution on 

the sensor.  

4.  Seal the jar. We usually use plastic wrap and rubber bands. 
5.  Start the program and allow some time (2 to 6 hours) for the air inside the sensor 

to reach the proper relative humidity level.  

6.  Go through the procedure for the first point of calibration. Type in the relative 

humidity for the salt you used, as determined from the table. 

 

4

7.  Repeat the procedure for the second calibration point using a different salt. Be 

sure to allow enough time for the Relative Humidity Sensor to adjust for the 
change in humidity. 

Even though none of the salts listed here is especially dangerous, use normal 

precautions with these chemicals. 
Here are the relative humidity readings to use when calibrating with salts. These 

numbers were taken from Hy-Cal Engineering IH-3602-L data sheets. 
 

 

All Data in %  

15°C 20°C 25°C 30°C 35°C 

Lithium 

Bromide 

6.86 6.61 6.37 6.16 5.97 

Lithium 

Chloride 

11.3 11.31 11.3 11.28 11.25 

Potassium Acetate 

23.40 23.11 22.51 21.61 - 

Magnesium Chloride 

33.3 

33.07 32.78 32.44 32.05 

Potassium Carbonate 

43.15 43.16 43.16 43.17 - 

Magnesium Nitrate 

55.87 

54.38 52.89 51.4  49.91 

Potassium Iodide 

70.98 

69.90 68.86 67.89 66.96 

Sodium 

Chloride 

75.61 75.47 75.29 75.09 74.87 

Ammonium Sulfate 

81.70 

81.34 80.99 80.63 80.27 

Potassium Chloride 

85.92 

85.11 84.34 83.62 82.95 

Potassium 

Nitrate 

95.41 94.62 93.58 92.31 90.79 

Warranty 

Vernier warrants this product to be free from defects in materials and workmanship for a 
period of five years from the date of shipment to the customer. This warranty does not 
cover damage to the product caused by abuse or improper use. 

 

 
 
 
 

 

Vernier Software & Technology 

13979 S.W. Millikan Way 

 Beaverton, OR 97005-2886 

Toll Free (888) 837-6437 

 (503) 277-2299 

 FAX (503) 277-2440  

[email protected] 

 www.vernier.com 

 
Rev 6/5/2012 
Logger

 Pro

, Logger Lite, Vernier LabQuest 2, Vernier LabQuest, Vernier LabQuest Mini, Vernier LabPro, Go! Link, 

Vernier EasyLink and other marks shown are our trademarks or registered trademarks in the United States. 
TI-Nspire, CBL 2 and CBL, TI-GRAPH LINK, and TI Connect are trademarks of Texas Instruments. 
All other marks not owned by us that appear herein are the property of their respective owners, who may or may not be 
affiliated with, connected to, or sponsored by us.  

 

Printed on recycled paper. 

Отзывы: