background image

 

 5

Determining the Wavelength(s) to Use in an Experiment   

When you conduct a Beer’s law lab or a kinetics lab, it is common to select one 

wavelength at which to follow the experiment. However, in Logger Pro 3.4.6 you 

may select as many wavelengths as you wish. There are two ways to select the 

wavelength or wavelengths. 
1. 

Perform a Full Spectrum Analysis of the Solution to Be Tested

 

Measure the full spectrum of a sample of solution and then click Configure 
Spectrum Data Collection (

). Select Abs 

vs.

 Concentration or Abs 

vs

. Time. 

The wavelength of maximum absorbance (

 max) will be automatically selected. 

2. 

Use a Sample of Solution to Determine the Peak Absorbance

 

This is a variation of the previous method. After calibrating the Spectrometer, 
place a sample of solution in the Spectrometer and then click Configure Spectrum 
Data Collection (

). Select Abs 

vs.

 Time or Abs 

vs.

 Concentration. The 

wavelength of maximum absorbance (

 max) will be automatically selected. If 

you don’t want to use the 

 max, click

 and select a wavelength on the 

graph or in the list of wavelengths. 

Selecting a Range of Wavelengths to Use in an Experiment 

You may wish to measure the absorbance or %T of a sample over a group of 

wavelengths rather than a single wavelength. There are two ways to select a group of 

wavelengths from the Configure Spectrum Data Collection dialog box.  

 

Select the wavelengths one at a time by checking the boxes in the Select 

Wavelength column. 

 

Place the cursor on the graph in the dialog box. Left click and drag across the 

region of wavelengths that you wish to analyze. Make sure to check the “Treat 

Contiguous Wavelengths as a Single Range” box. 

Measurement 

You can set up the spectrometer to measure intensity, absorbance, or 

% transmittance. Choose Change Units 

 Spectrometer from the Experiment menu. 

Click on the unit of choice from the list. 

Using an Ocean Optics Spectrometer with a LabQuest 2 or 
Original LabQuest 

1.  Use the USB cable to connect the spectrometer to a LabQuest 2 or original 

LabQuest. Allow the spectrometer to warm up for a few minutes. 

2.  Turn on the LabQuest 2 or LabQuest. The LabQuest app will launch 

automatically and the meter screen will be displayed. 

Calibrate the Spectrometer 

1.  Fill a cuvette about ¾ full with distilled water and place it in the spectrometer. 

Align the cuvette so a clear side of the cuvette is facing the light source.  

 

 6

2. Choose Calibrate 

 USB:Spectrometer from the Sensors menu. At the prompt, 

select Finish Calibration. After the message “Calibration Completed” appears, 
select 

Measure the Absorbance Spectrum of an Aqueous Sample 
(Absorbance 

vs.

 Wavelength) 

1.  Fill a cuvette about ¾ full of the solution to be tested and place it in the 

spectrometer. 

2.  Start the data collection. Tap the red Stop button to end data collection. 
3. 

Note:

 The wavelength of maximum absorbance (

λ

 max) is automatically selected. 

This 

λ

 max will be used for any subsequent data collection, such as a Beer’s Law 

experiment (abs 

vs.

 conc.) or a kinetics experiment (abs. 

vs

. time). If you wish to 

choose another wavelength, you can tap on the graph to select a new wavelength 
or you can used the arrow keys on the keypad to move the cursor to a new 
wavelength. 

Conduct a Beer’s Law Experiment (Absorbance 

vs.

 Concentration) 

1.  Measure an absorbance spectrum as described above. On the Meter screen, tap 

Mode. Change the mode to Events with Entry. 

2.  Enter the Name (Concentration) and Units (mol/L). Select OK. 
3.  A message will appear warning you to either save or discard the full spectrum 

run. Make your choice and proceed with the data collection. 

4.  Place your first Beer’s law standard solution in the spectrometer. Start the data 

collection. After the absorbance reading stabilizes, tap Keep. Enter the 
concentration of the solution and select OK. 

5.  Place your second standard sample in the spectrometer. After the absorbance 

readings stabilize, tap Keep. Enter the concentration of the second sample and 
select OK. 

6.  Repeat Step 5 for the remaining standard samples. After you have tested the final 

standard, tap the red Stop button to end the data collection. 

7.  To calculate a best fit line equation for your standards, choose Curve Fit from the 

Analyze menu. Select Linear for the Fit Equation, and then select OK. The graph 
screen will appear again with the linear regression equation displayed. 

8.  Place a cuvette containing an unknown sample of solution in the spectrometer. 

Tap the Meter tab and write down the displayed absorbance value. Tap the graph 
tab and trace the linear regression equation to determine the concentration of the 
unknown. 

Conduct a Kinetics Experiment (Absorbance 

vs

. Time) 

1.  Measure an absorbance spectrum as described above. On the Meter screen, tap 

Mode. Change the data-collection mode to Time Based. 

2.  You can change the rate, interval, and/or length of time of data collection, if 

desired. Select OK when you are ready to proceed. 

Отзывы: