Vernier Go Wireless GW-PH Скачать руководство пользователя страница 2

Specifications 

 

pH Electrode 

pH range 

0–14 

Temperature range 

5 to 80°C (readings not compensated) 

Accuracy 

±0.5 pH units (factory calibration) 
±0.1 pH units (user calibration) 

Type Single 

junction 

Sealed, gel-filled, Ag/AgCl  

Response time 

2 s (to 90% of full reading in buffer) 

Shaft diameter 

12 mm OD 

Shaft material 

Polycarbonate plastic 

 

Electrode Amplifier 

Radio 

2.4 GHz Bluetooth 4.0, Low Energy 

Battery 

100 mA Li-Poly 

Battery life (continuously on,  
   1 sample per second) 

3 days 

Input range 

±1000 mV 

Temperature range 

–10 to 45ºC 

Maximum wireless range 

30 m unobstructed 

 

How Go Wireless pH Works 

The electrode amplifier handle is a circuit that allows the standard combination pH 
electrode to be monitored by a lab interface wirelessly.  
The Go Wireless pH will produce a voltage of approximately 0.0 volts in a pH 7 
buffer. The voltage will increase by about 60 mV for every pH number decrease. 
The voltage will decrease by about 60 mV/pH number as the pH increases. It should 
read approximately 160 mV in pH 4 buffer and 

170 mV in pH 10 buffer.  

The Vernier gel-filled pH Electrode is designed to make measurements in the pH 
range of 0 to 14. The gel-filled reference half cell is sealed; it cannot be refilled.  
 

Go Wireless Electrode Amplifier has one button and three LEDs. 

Button 

 

Press the button once to turn on Go Wireless Electrode Amplifier. A red LED 
indicator flashes when the unit is on.  

 

Press and hold the button for more than three seconds to put Go Wireless 
Electrode Amplifier into sleep mode. The red LED indicator stops flashing when 
sleeping and LED light is off. 

 

Press and hold the button for more than eight seconds to reset the system. This 
should not be necessary in normal operation.  

 

Blue LED 

Indicates charging status when Go Wireless Electrode Amplifier is connected to the 
Charging Cable or Charging Station. 

 

On  – charging in progress 

 

Off – charging is complete 

 

 

Red LED Flashing 

Indicates Go Wireless Electrode Amplifier is awake and ready to connect. 

Green LED Flashing 

Indicates Go Wireless Electrode Amplifier is connected and taking data. 

Red and Green LED Off  

Indicates Go Wireless Electrode Amplifier is in the lowest power sleep mode. 
 

Collecting Data  

Ensure that the BNC connection between the Go Wireless Electrode Amplifier and 
the pH Electrode is secure.  
 

 

 
To prepare the electrode to make pH measurements, remove the storage bottle from 
the electrode by first unscrewing the lid and then removing the bottle and lid. 
Thoroughly rinse the lower section of the probe, especially around the bulb-shaped 
tip, using distilled or deionized water. 

Note:

 Do not completely submerge the sensor. 

The handle is not waterproof. 

 

Collecting Data with Go Wireless App  

1. Press the power button on Go Wireless Electrode Amplifier to turn it on. The 

LED on the sensor will flash red. 

2. Launch the app. Select your Go Wireless pH from the list of available sensors. 

When the sensor is paired with the software, the sensor LED will flash green.  

 

Note: 

To collect data in mV, tap on the radio icon in the upper right corner. Under 

Units, select mV. 

3. Place the sensor tip into a beaker containing your sample. It is recommended that 

you use a utility clamp or electrode support to fasten the pH Sensor to a ring 
stand, as shown on next page. 

Содержание Go Wireless GW-PH

Страница 1: ...harging status A blue light indicates it is still charging When the light goes off it is finished charging 2 Supported Devices Go Wireless pH can be used with supported mobile devices that have Bluetooth Smart support and one of our apps iPad 3rd generation or newer iPad mini and iPad Air iPhone 4S or newer iPod touch 5th generation or newer LabQuest 2 model LQ2 LE see www vernier com til 3085 For...

Страница 2: ...cator stops flashing when sleeping and LED light is off Press and hold the button for more than eight seconds to reset the system This should not be necessary in normal operation Blue LED Indicates charging status when Go Wireless Electrode Amplifier is connected to the Charging Cable or Charging Station On charging in progress Off charging is complete 4 Red LED Flashing Indicates Go Wireless Elec...

Страница 3: ... Go Wireless pH rinse the tip of the electrode with distilled water pat dry and place the electrode in the first buffer solution so the tip is immersed 5 Once the voltage reading has stabilized tap Add in the box under the appropriate buffer pH Note If you are using a buffer not shown tap on the buffer value to change 6 Repeat this process for two or three buffer solutions Your stored calibration ...

Страница 4: ...asily be removed using a mild detergent solution Mold growth in the storage solution can be inhibited by adding a buffer preservative The Go Wireless pH is designed to be used in aqueous solutions The polycarbonate body of the sensor can be damaged by many organic solvents In addition do not use the sensor in solutions containing perchlorates silver ions sulfide ions biological 8 samples with high...

Страница 5: ...compliance could void your authority to operate the equipment IC Statement This device complies with Industry Canada license exempt RSS standard s Operation is subject to the following two conditions 1 this device may not cause interference and 2 this device must accept any interference including interference that may cause undesired operation of the device Industry Canada Class B This digital app...

Отзывы: