Vernier GC-MINI Скачать руководство пользователя страница 7

 

13

Collecting a Sample 

  1. Remove the cap from the sample container and rinse the syringe with the 

solution to be injected, following the procedure outlined in the section 

Rinsing 

the Syringe

  2. Submerge the tip of the needle in the solution to be injected and slowly draw 

the plunger back by an amount larger than the intended injection volume. This 
excess volume will typically be somewhere around 0.1 µL. (Example: If you 
are attempting a 0.2 µL injection, you should withdraw between the 0.25 and 
0.3 µL.) 

  3. Remove the syringe tip from the solution and replace the cap on the sample 

container. 

  4. Holding the syringe vertically at eye level, carefully depress the plunger until 

the syringe holds only the desired injection volume. 

  5. Carefully wipe the syringe needle with a fresh lint-free tissue, making sure the 

tissue does not absorb any sample from inside the needle. 

  6. When you are satisfied with the volume of your sample, proceed to injection. If 

not, carefully eject the syringe contents into the appropriate waste receptacle, 
wipe the needle with a fresh lint-free tissue, and begin again at Step 2.  

Injection into the Gas Chromatograph 

  1. Wait for the Mini GC’s status indicator light to turn green and the message 

“Ready for Injection” to appear before collecting your sample into the syringe. 

  2. As soon as your sample is loaded into the syringe, grip the syringe by the 

flange and pinch the needle about halfway along its length. 

  3. Aim the needle into the center of the brass needle guide at the top of the gas 

chromatograph and slowly push the needle into the injection port, 

being careful 

not to move the plunger

. If the needle should catch, immediately stop pushing 

and turn the syringe a quarter turn, then resume pressure. 

  4. As soon as the syringe is fully inserted against the needle guard, quickly 

depress the plunger and simultaneously click Collect in the data-collection 
software. 

  5. Immediately pull the syringe out of the gas chromatograph. 

Cleaning the Syringe 

  1. Never force the plunger.

 

  2. Clean the syringe following the procedure described in 

Rinsing the Syringe

 

using an appropriate solvent. Choose a solvent that will dissolve the 
contaminating substances. Common choices include acetone, methanol, 
methylene chloride, and acetonitrile, but other solvents may used.  
Non-alkaline, non-phosphate, non-detergent solvents are preferred. Do not 
soak the syringe in any solvent.

 

  3. If the plunger is catching during injection, it may need to be flushed with 

multiple cleaning solvents. 

 

 

 

14

Troubleshooting 

Syringe plunger won’t fully depress. 

  1. Follow the procedure for cleaning the syringe 
  2. Inspect the plunger. If it is bent and cannot be straightened, dispose of the 

syringe according to your sharps disposal procedure. 

Appendix D 

Relationship of Pressure and Column Flow Rate 

If you are accustomed to setting 

flow rate

 rather than 

column pressure

 values in a 

temperature-pressure profile, here is an equation (and a table of typical values) 

representing the nearly linear relationship between these two variables:  

Flow Rate (mL/min) = 0.54

 

 

Pressure (kPa) 

 

Pressure (kPa) 

Flowrate (mL/min) 

2.5 1.4 
5.0 2.7 
7.5 4.1 

10.0 5.4 
12.5 6.8 
15.0 8.1 
17.5 9.5 
20.0 10.8 

This relationship was determined by measuring the flow rate of the exiting gas (air) 

when the column was heated at a constant 45

C temperature at various pressures. 

Warranty 

Vernier warrants this product to be free from defects in materials and workmanship 
for a period of two years from the date of shipment to the customer. Consumable 
items, such as the MEMS sensor, the Restek GC column, and syringes are excluded 
from the warranty. Damage to any part of the Vernier Mini GC product because of 
misuse, misapplication, neglect, alteration, accident, and/or operation contrary to 
design or damage resulting from natural causes, is not covered by this warranty.

 

 

Note:

 This equipment has been tested and found to comply with the limits for a 

Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are 

designed to provide reasonable protection against harmful interference when the 

equipment is operated in a commercial environment. This equipment generates, uses, 

and can radiate radio frequency energy and, if not installed and used in accordance 

with the instruction manual, may cause harmful interference to radio 

communications. Operation of this equipment in a residential area is likely to cause 

harmful interference in which case the user will be required to correct the 

interference at his own expense. 
 

Содержание GC-MINI

Страница 1: ...priate personal safety equipment and use the instrument in accordance with the instructions in this manual to prevent any injury 2 Caution The Vernier Mini GC gas chromatograph is designed to analyze...

Страница 2: ...eedle or bend the plunger Never pull the plunger back more than 50 of its total volume Be careful not to bend the plunger as you press it down a Depress the plunger fully b Submerge the tip of the syr...

Страница 3: ...he Vernier Mini GC The power input jack power switch USB connection to the computer or LabQuest and column ventilation grill are shown in this view Principles of Operation The Vernier Mini GC is desig...

Страница 4: ...he power supply into the power socket on the side of the gas chromatograph insert the 120 VAC plug into an outlet Connect the mini jack of the USB cable to the instrument and the standard USB jack int...

Страница 5: ...ced by unscrewing counterclockwise the septum retainer nut using a coin or a large screwdriver Note Two additional Septa were shipped with your new Mini GC in a small Ziploc bag With the retainer nut...

Страница 6: ...ill destroy the syringe Rotating the syringe will alleviate this problem 12 8 When wiping the needle of the syringe pinch a lint free wipe around the needle next to the barrel of the syringe and wipe...

Страница 7: ...ssolve the contaminating substances Common choices include acetone methanol methylene chloride and acetonitrile but other solvents may used Non alkaline non phosphate non detergent solvents are prefer...

Страница 8: ...uct contact your local city office your disposal service or the place where you purchased it The symbol shown here indicates that this product must not be disposed of in a standard waste container 16...

Отзывы: