background image

 
 

Logger 

Pro

 3 (computers) 

1.  Set up the equipment and start Logger 

Pro

 3. 

2.  Open the Drop Counter folder located in the Probes 

& Sensors folder. If one of the file names seems 
appropriate, open it and proceed normally. If none 
of these file names seem appropriate, open the file 
named “Drop Counter-pH” and proceed with 
Step 3. 

3.  Choose Show Sensors from the Experiment menu. 

4.  Click on the picture of the pH Sensor in Channel 1 and select Choose Sensor 

from the pop-up menu. 

5.  Choose the sensor you will be using from the drop-down list. 

6.  Click on the picture of your sensor in Channel 1 and select Calibrate from the 

pop-up menu. 

7.  Calibrate the sensor using one of these methods. 

 

If you wish to perform a new calibration, proceed with the calibration 
instructions outlined earlier in this manual. 

 

If you wish to manually enter a calibration, choose Equation/Options from the 
Live Calibration pull-down list. Enter the appropriate slope and intercept from 
a previous calibration or from the chart below. 

DataDrop (TI graphing calculators) and Data Pro (Palm OS handhelds) 

1.  Connect a Vernier auto-ID sensor, then press the Clear button on a TI graphing 

calculator, or choose New from the Data Pro menu with a Palm OS handheld. In 
both cases, the sensor you have connected should automatically be identified, and 
a calibration loaded.  

2.  If you use a non-auto-ID sensor, or if the calibration that is loaded by default is 

not the unit you wish to use, another option is to manually select the new sensor, 
and/or manually enter a new calibration value. To do this, go to Setup in the 
DataDrop or Data Pro programs, and choose a new sensor, or choose to manually 
enter a new calibration (you can enter a slope and intercept value for the sensor 
you are using from the chart below).  

 

 Slope 

Intercept 

Conductivity (0–200 µS/cm) 

65.70 

Conductivity (0–2000 µS/cm) 

847.2 

Conductivity (0–20000 µS/cm) 

7819 

ISE (mV) 

137.55 

–168.23 

Electrode Amplifier (mV) 

223.21 

–390.63 

pH Amplifier (mV) 

223.21 

–390.63 

pH Amplifier (pH units) 

–3.838 

13.72 

 

 
 

10

LabQuest 

1.  Set up the equipment and start LabQuest App. 

2.  If your sensor does not auto-ID, choose Sensor Setup from the Sensors menu. 
3.  Open the drop down menu next to the Channel in which the sensor is connected. 

Select the sensor you are using from the list. 

4.  Select OK to continue. 

Vernier Drop Counter Specifications 

 

Detector rise time: <70 ns 

 

Detector fall time: <70 ns 

 

Power requirements: 55 mA at 5 VDC 

 

Infrared source: Peak at 890 nm 

Limited Warranty 

Vernier warrants this product to be free from defects in materials and workmanship 
for a period of five years from the date of shipment to the customer. This warranty 
does not cover damage to the product caused by abuse or improper use. 

 