Troubleshooting
Here are some tips for best data-collection practices:
Allow the Colorimeter to warm up for about 5 minutes before calibrating.
Fill a cuvette two-thirds to three-fourths full with liquid, including the
calibration blank, so that the light travels through the liquid reliably.
After filling a cuvette with liquid, seal the cuvette with a cap to prevent spills.
Make sure to place a cuvette in the Colorimeter so the path of the light source
travels through the clear sides of the cuvette. An arrow to the right of the
cuvette slot shows the light path.
For best results, use one cuvette to make all your measurements for a given
experiment.
If you calibrate a Colorimeter and then change the wavelength, calibrate the
Colorimeter again to ensure the proper identification of the new wavelength.
For best results, a sample's absorbance or transmittance values should fall within
these ranges:
percent transmittance:10–90%
absorbance: 0.05–1.0
We have found that Beer’s law experiment results begin to lose their linearity at
absorbance values above 1.0 (percent transmittance values less than 10%). If you
have a solution that transmits such a low level of light, consider diluting the
solution so that it falls within this range.
Repair Information
If you have watched the related product video(s), followed the troubleshooting
steps, and are still having trouble with your Colorimeter, contact Vernier
Technical Support at [email protected] or call 888-837-6437. Support
specialists will work with you to determine if the unit needs to be sent in for
repair. At that time, a Return Merchandise Authorization (RMA) number will be
issued and instructions will be communicated on how to return the unit for repair.
Accessories/Replacements
Item
Order Code
Replacement cuvettes (package of 100 with 20 lids)
CUV
Replacement cuvette lids (package of 100)
CUV-LID
Cuvette rack, holds 10 cuvettes
CUV-RACK
Warranty
Vernier warrants this product to be free from defects in materials and workmanship
for a period of five years from the date of shipment to the customer. This warranty
does not cover damage to the product caused by abuse or improper use. This
warranty covers educational institutions only.
This warranty does not cover consumables, such as the cuvettes and lids.
3
Vernier Software & Technology
13979 SW Millikan Way • Beaverton, OR 97005-2886
Toll Free (888) 837-6437 • (503) 277-2299 • FAX (503) 277-2440
[email protected] • www.vernier.com
Rev. 2/8/16
Logger
Pro,
Logger Lite, Vernier LabQuest 2, LabQuest Mini, and other marks shown are our trademarks or
registered trademarks in the United States.
All other marks not owned by us that appear herein are the property of their respective owners, who may or may
not be affiliated with, connected to, or sponsored by us.
4