NeuroTrac
®
Continence Operation Manual
3
* This unit must be used with the guidance of a Physiotherapist or Doctor.
* Type BF equipment, Continuous Operation.
* Do not insert lead wires into a mains power supply.
* Do not immerse unit into water or any other substance.
* The unit is not protected from the ingress of shower of rain if used outside
the carrying case.
* Do not use the NeuroTrac® Continence unit in the presence of a
flammable anaesthetic gas mixture.
* Please use 9 Volt alkaline battery (IEC code: 6LR61, size: PP3). If using
rechargeable 9 Volt battery, be sure to use Nickel Metal Hydride type
batteries (IEC Code 6HR61).
We advise not to use Ni-Cad rechargeable batteries.
Caution: Do not use lithium batteries unless they comply with
IEC60086-4.
* Patient Electrodes are for single patient use only.
* Keep out of reach of children.
* Do not use this stimulator on your facial area unless you are under strict
guidance from a qualified Clinician.
* Application of electrodes near the thorax may increase the risk of cardiac
fibrillation.
* Operation in close proximity to shortwave or microwave therapy
equipment may produce instability in the stimulator output.
* Simultaneous connection of a patient to high frequency surgical equipment
may result in burns at the site of the stimulator electrodes and possible
damage to the stimulator.
* This device can deliver current densities in excess of 2mA/cm2 when used
at a high intensity with small electrodes.
* No modification of this equipment is allowed.
Warnings