NeuroTrac

®

Continence Operation Manual

1

DUAL CHANNEL STIM UNIT

Operators Manual

Visit our website: www.veritymedical.co.uk  

for detailed application protocols

 

English

 

 

NeuroTrac  Continence

 

®

Содержание NeuroTrac Continence

Страница 1: ...NeuroTrac Continence Operation Manual 1 DUAL CHANNEL STIM UNIT Operators Manual Visit our website www veritymedical co uk for detailed application protocols English NeuroTrac Continence ...

Страница 2: ...r when you report a technical fault or claim a warranty return Manufacturer s serial number of the unit Present it together with LOT number when you report a technical fault or claim a warranty return Name and address of Manufacturer Date of manufacture Conformity indication with the essential health and safety requirements set out in European Directives 0088 Notified body identification LRQA Ltd ...

Страница 3: ...rechargeable batteries Caution Do not use lithium batteries unless they comply with IEC60086 4 Patient Electrodes are for single patient use only Keep out of reach of children Do not use this stimulator on your facial area unless you are under strict guidance from a qualified Clinician Application of electrodes near the thorax may increase the risk of cardiac fibrillation Operation in close proxim...

Страница 4: ... 6 Description of STIM unit functions 7 Quick start instructions 8 9 Lock button 10 Setting up customised programmes 10 Continence treatment programmes 12 Electrode types tips 13 Care maintenance accessories and disposal 14 Indications for use 15 Technical specifications 16 Troubleshooting 17 Warranty 18 Clinical references 19 Contents ...

Страница 5: ...roTrac Continence is a dual channel device combining several treatment programmes into one unit Neuromuscular stimulation is increasingly understood by therapists and doctors There is a better understanding of the mechanisms which exist between nerves and muscles that makes it possible to stimulate the neuromuscular system with precise electrical signals The NeuroTrac Continence offers precision g...

Страница 6: ...otentially dangerous equipment Do not place electrodes Over carotid sinus nerves Over larynx or trachea Inside mouth Over the area of the heart unless so advised by your doctor On your facial area unless under strict guidance from a qualified clinician Do not apply stimulation across or through the head directly on the eyes covering the mouth on the front of the neck especially the carotid sinus o...

Страница 7: ...res customised programme and returns to the home position LCD Display Channel A Channel B SET to go to custom programme settings ESC to stop programme or escape from settings to start programme Select Programme Lock Button Lead Wire and Pin Front Back Concealed button locks the selected programme and records the pa tient s home compliance Use a metal pin of the lead wire to press the concealed but...

Страница 8: ...he lead wire It does not matter which way the red black cables are connected Some probes have a long wire built in with the direct plug to the unit 5 Make sure the vaginal or rectal electrode is clean before use 6 Insert the vaginal or rectal electrode using any electrically conductive lubricating gel or any other water based lubricating gel recommended for maximum performance A typical vaginal el...

Страница 9: ...ntly in mild soapy water rinse and make sure the probe is completely dry before returning to storage in the plastic bag Low Battery Indicator When the battery power is low the low battery indicator will appear on the screen shown in the diagram on page 7 When the battery indicator shows one bar replace the battery Disconnection Indicator When the probe becomes disconnected or when the lead wires d...

Страница 10: ... a thin rod end of the lead wire pin meets this requirement until a single beep is heard Now the LCD will display the average mA used on each channel and the total hours the unit has been in use as shown in the diagram To return to normal unlocked operation simply press ESC Setting up the Customised Programme PC4 Continuous Select PC4 by pressing the PRG button on the front panel 1 Press and hold ...

Страница 11: ...en press the or button to adjust the rest period 2 99 seconds 6 Press the SET button again and the RMU Ramp up symbol will be displayed then press the or button to adjust the ramp up period from 0 1 9 9 seconds 7 Press the SET button again and RMD Ramp down symbol will be displayed then press the or button to adjust the ramp down period from 0 1 9 9 seconds 8 Press the SET button again and SYN Syn...

Страница 12: ...s 30 Hz for 4 min 40 Hz for 3 min 200 µs ramp up time 0 7 s work time 4 s rest time 4 s P08 PELVIC FLOOR WORK OUT 60 min Sequential 20 Hz for 3 min 3 Hz for 5 min 10 Hz for 15 min 20 Hz for 15 min 250 µs ramp up time 0 8 s 30 Hz for 5 min 40 Hz for 5 min 200 µs ramp up time 0 6 s 10Hz for 12 min 250µs ramp up time 0 8 s Work time 4 s Rest time 4 s P09 BUILDING UP ENDURANCE P09 20 250 0 8 5 5 20 CU...

Страница 13: ...e If this does not work try cleansing the skin with a swab impregnated with alcohol Clip away hairy skin using scissors The electrodes conductive material is water based If it becomes saturated e g from perspiration it will lose its adhesive qualities After use leave the electrodes face up overnight to dry out replace on plastic film in the morning At some point the electrodes will become dry Mois...

Страница 14: ...y for any discharge from the battery Remove battery completely from unit if not in use for any extended period of time Low battery indicator of 6 9 volts shown on LCD display When flashing change battery for a new one Please use 9 Volt alkaline battery IEC code 6LR61 size PP3 Battery disposal must be disposed of in compliance with national regulatory requirements Lead Wires The lead wires should b...

Страница 15: ... they drop onto the floor debris will adhere to conductive gel rendering the electrodes ineffective Electrode life can be considerably reduced by The type and condition of the skin Deep seated moisturisers or make up For the Best Results Before each use cleanse the skin After each use stick the pads on the shiny insert card and store in a cool and dry place For troubleshooting see page 17 Caution ...

Страница 16: ...mer can be set up to 12 h 59 min In PC1 PC3 the max time is 1h 30 min 8 Ramp up time 0 3 9 9 seconds 9 Battery 9 Volt alkaline battery IEC code 6LR61 size PP3 Expected aver age battery life of standard 800 mAh alkaline 32 hours It is recommended to use battery type 6LR61 10 Low Battery Indicator If the battery goes below 6 9 volts 0 2 volts the battery symbol will flash on off once every second 11...

Страница 17: ...osition is to sit down and lean back slightly or lie down 5 If you think the probe itself is not working wash it and hold it using your first finger and thumb or elbow crook to make a connection across the electrode plates Connect it to the stimulator as normal Increase mA and if the probe is functioning correctly you will feel the stimulation mildly tingling in your hand or elbow crook This prove...

Страница 18: ...with the unit or normal wear and tear Any evidence of tampering will nullify this war ranty Customer Service Please contact your distributor for any customer service enquiries including the warranty returns Your invoice of purchase and or the rear cover of this manual should state the name and the contact details of your distributor For assistance if needed in setting up using or maintaining the u...

Страница 19: ...ion Manual 19 Clinical References Please go to our website for the latest clinical protocols http www veritymedical co uk Protocols Please contact us for any clinical references of NeuroTrac Continence enquiries veritymedical com ...

Страница 20: ...NeuroTrac Continence Operation Manual 20 Distributor Not for sale or use in the USA Continence Accessory control information NeuroTrac Continence manual English ECS400A OM EN21 15 09 17 ...