background image

8

7

To enter the configuration mode, the user should send a
predetermined command from the host to VH2R (letter ‘c’
in lower case). The power down timer will be extended upon
receiving any character from the host system. When the
timer expires, the unit will power down. All configuration
settings are stored in non-volatile memory immediately
after they are entered. To exit the configuration mode, the
user can either use a special command from PC – letter ‘q’
in lower case (the VH2R will go into Idle mode), or press the
F1 button (the VH2R will enter scan mode).

An example of configuration menu screen:

VeriChip H2 Reader CONFIG MODE 
m - BT Master
s - BT Slave
u - USB
r - Ricoh G3

@

- BT MAC

# - Unlock Code
? - Unlock Required
! - Unlock Not Required
q - Quit

VeriMed H2 Reader Communication

The unit has two communication devices – USB and Bluetooth.
Only one communication mode can be used at any time. 

Both communication channels are bi-directional. When a
passive VeriMed Microtransponder is acquired by the
VH2R Reader, the ID number is sent to the host computer.
The host computer can also send certain information back
to the VH2R Reader. This feature is used to change current
configurations of the VH2R Reader, including possible a
switch between its communication modes (from USB to
BT and back) and utilizing the Authorization Code option.

VeriChip H2™ Reader Operation in USB Mode

To operate in this mode the user should perform the
following actions:
• Connect the USB cable from the host computer to the

VH2R Reader. The USB port is located in the rear of the
case. Lift the protective cover and insert the cable.

• To perform the scan, the user should press the F1

button once again. The VH2R will turn on the LF (low
frequency) interrogation field and try to acquire a
Microtransponder. The field will stay on for 12 seconds
or until a valid Microtransponder is acquired. If a
Microtransponder has not been acquired, the VH2R will
go into the idle mode. If a Microtransponder has been
acquired, its ID number will be displayed on VH2R LCD
and sent to the host computer via the USB interface.
After this the VH2R will enter its idle mode and be ready
for another scan session or power down after 5
minutes. The unit can be powered down from this state
by pressing the two smaller buttons (F2 and F3)
simultaneously.

Note: 

Operating in the USB mode, the unit does not monitor

the status of communication link. That is, if the USB cable
becomes disconnected, the user will not be notified. If the
cable was not connected to the host system, the latter may
not be able to open the serial port. Disconnecting the cable
during the communication session may produce an error on
the host system.

VeriChip H2 Reader Operation 
in Bluetooth Master Mode

In the Master mode the VH2R performs the discovery and
then dials the host computer system acting as a Bluetooth
(BT) master. As a consequence, no action is required on the

Содержание VeriChip H2

Страница 1: ...O P E R A T O R S M A N U A L AN APPLIED DIGITAL COMPANY...

Страница 2: ...Press Button F1 and release to power on the Reader 2 The Reader will enter a self diagnostic mode displaying the following on the LCD screen a Initializing b VeriChip H2 Reader BT Master Rev x xx c A...

Страница 3: ...ally enter the ID number Turning the Reader Off The Reader can be turned off manually at any time by simultaneously pressing and holding down the F2 and F3 buttons If the Authorization Code is enabled...

Страница 4: ...al should be set to 9600 Baud 8 data bits one stop bit no parity 8 N 1 To enter the configuration mode the user should send a predetermined command from the host to VH2R letter c in lower case The pow...

Страница 5: ...tion in USB Mode To operate in this mode the user should perform the following actions Connect the USB cable from the host computer to the VH2R Reader The USB port is located in the rear of the case L...

Страница 6: ...l not require entering the passkey unless the devices become unpaired The operation is performed as follows Tooperateinthismodetheusershouldperformthefollowingactions Bring the VH2R and host system wi...

Страница 7: ...the electromagnetic environment specified below The customer or the user of the VeriChip H2 Reader should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance le...

Страница 8: ...ty continued on next page Table 204 Guidance and manufacturer s declaration electromagnetic immunity continuation Immunity Test IEC 60601 test level Compliance level Recommended separation distance Ra...

Страница 9: ...ing the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and...

Страница 10: ...in which it will be used Warning The VeriChip H2 Reader generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful...

Страница 11: ...measures Reorient or relocate the receiving antenna Increasetheseparationbetweentheequipmentandthereceiver Consultthedealeroranexperiencedradio TVtechnicianforhelp Modifications Any changes or modifi...

Отзывы: