ELECTROMAGNETIC COMPATIBILITY
GUIDANCE AND MANUFACTURE’S DECLARATION-ELECTROMAGNETIC IMMUNITY
FOR ALL EQUIPMENT AND SYSTEMS
Guidance and manufacture’s declaration-electromagnetic immunity
The Pulse Oximeter is tended for use in the electromagnetic environment specified below. The user of
the Pulse Oximeter should assure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD) IEC 61000-4-2
Power frequency
(50Hz) magnetic field
IEC 61000-4-8
IEC60601 test level
±6KV contact
±8KV air
3A/m
Electromagnetic environment-
guidance
Floors should be wood,
concrete or ceramic tile. If floor
are covered with synthetic
material, the relative humidity
should be at least 30%.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment
Compliance level
±6KV contact
±8KV air
3A/m