![VERDER CARBOLITE GERO Скачать руководство пользователя страница 20](http://html.mh-extra.com/html/verder/carbolite-gero/carbolite-gero_installation-operation-and-maintenance-instructions_1003717020.webp)
l
When the gas is flowing, turn the adjustment knob on
the flowmeter to set the desired gas flow rate.
Note:
If using the Advanced gas box, the gas will only
flow when the solenoid valve is activated.
l
Check that the correct gas flow rates are set. The
reading should be taken from the top of the float so
that the flat edge is level with the increments on the
gas scale.
Note:
To minimise the increase of back pressure, always increase the
pipe diameter of any room vent lines when connecting to a product
exhaust outlet. For example, 6mm outlet connections should be increased
to a diameter of 10mm or greater for room lines between 5 - 10 metres
in length.
Note:
Always maintain clean gas lines. Regularly inspect vent lines
connected to the exhaust outlet as any build up of debris may increase the
back pressure of the vessel/ work tube.
6.1
Gases and Flow Rates
Gas
Flow Rate Range (litres/minute)
Air
1 - 10
Nitrogen (N
2
)
1 - 10
Argon (Ar)
1 - 8
95% Nitrogen (N
2
) / Hydrogen 5% (H
2
)
1.2 - 11
20