8.0 Fault Analysis
8.0
Fault Analysis
A.
Furnace Does Not Heat Up
1.
The temperature
controller is OFF
No power from the supply
Check the fuses in the
supply line
There are no lights
glowing on the controller
The controller may be faulty
or not receiving a supply due
to a faulty switch or a wiring
fault
2.
The temperature
controller is ON
The controller shows a
very high temperature or
a code such as S.br
The thermocouple has
broken or has a wiring fault
The controller shows a
low temperature
The SSR could be failing to
switch on due to internal
failure, faulty logic wiring
faulty which could have
overloaded the SSR
34