6.0
Maintenance
6.1
General Maintenance
Preventive rather than reactive maintenance is recommended. The type and frequency
depends on the product use; the following are recommended.
6.2
Maintenance Schedule
CUSTOMER
QUALIFIED PERSONNEL
DANGER! ELECTRIC SHOCK
. Risk of fatal injury. Only electrically
qualified personnel should attempt these maintenance procedures.
Maintenance
Procedure
Method
Frequency
Daily Weekly Monthly
Bi-
Annually
Annually
Safety
Over-Temperature Safety Circuit
(if fitted)
Set an over-temperature setpoint lower
than the displayed temperature and
check for an over-temperature alarm as
detailed in the relevant controller manual
Over-Temperature Safety Circuit
(if fitted)
Electrical measurement
Door Seal
Visual inspection - check for splits and
make sure it seals
Door Seal
Replacement
Electrical Safety (external)
Visual check of external cables and plugs
Electrical Safety (internal)
Physically check all connections and
cleaning of the power plate area
Gas Safety - ISO model (external)
Carry out the leak test as detailed in this
manual.
Gas Safety – ISO model (internal)
If the leak test has failed and the door is
making a good seal then an internal leak
test is needed.
Air System Integrity (ASTM model)
Carry out the leak test as detailed in this
manual.
Function
Temperature Calibration
Tested using certified equipment, fre-
quency dependent on the standard
required
Operational Check
Check that all functions are working nor-
mally