When heating large objects, in particular poor conductors, avoid shielding the
thermocouple from the heating elements. The thermocouple is intended to sense the
temperature near the heating elements. However, if a large object is placed in the
chamber it may record the average temperature of the object and the elements, this
can lead to overheating of the elements. Allow large objects to gain heat at a lower
temperature and then reset the controller to a temperature close to the desired
maximum, or heat using a slowly controlled ramp rate. For more information refer to
the controller instructions.
The product's elements are very susceptible to mechanical shock. Take great care when
loading or unloading the chamber. If it is necessary to load or unload work at elevated
temperatures, keep the door open for as short a period as possible. The insulation cools
quickly and may crack though thermal shock.
For improved insulation and element life it is recommended to heat and cool at a slow
ramp rate e.g. 5 °C/minute, and to avoid opening the door at high temperatures.
On first installing the elements and on subsequent element replacement, operate the
product at 1500 °C for an hour to create a protective glaze on the element surface.
47
6.0 Operation
Содержание CARBOLITE GERO HTF 17/10
Страница 7: ...Caution Double Pole Neutral Fusing 7 1 0 Symbols and Warnings ...
Страница 38: ...4 0 3508 Dual Loop Cascade Power Limit View 38 ...
Страница 60: ...8 0 Repairs and Replacements 60 ...
Страница 65: ...G Grey GR Y Green Yellow 65 10 0 Wiring Diagrams ...
Страница 67: ...Please refer to the rating label for product specific information 67 11 0 Fuses and Power Settings ...
Страница 69: ...Notes Service Record Engineer Name Date Record of Work ...